CDS/R/4K DILUENT/SHEATH, 20L
Report
- Report Number
- 2919069-2007-00006
- Event Type
- Malfunction
- Date Received
- January 4, 2007
- Date of Event
- November 9, 2006
- Report Date
- December 15, 2006
- Manufacturer
- ABBOTT DIAGNOSTICS DIVISION/CELLDYN
- Product Code
- GKL
- PMA / PMN Number
- K961439
- Removal / Correction Number
- 2919069-1/4/07-001-C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NV, US
- Reporter Occupation
- OTHER
Narratives
USE OF DILUENT SHEATH LOT 42063I2 CAN CAUSE ERRORS IN RESULTS WHEN USED ON THE CELL-DYN RUBY, CELL-DYN SAPPHIRE AND CELL-DYN 4000. RESULTS THAT MAY BE IMPACTED INCLUDE RBC, MCV, RDW, PLATELETS AND MPV. THE IMPACT TO RESULTS VARIES BY INSTRUMENT TYPE AS WELL AS THE MATERIAL TESTED, I.E., CONTROL, CONTROL TYPE OR WHOLE BLOOD. RESULTS THAT MAY BE IMPACTED INCLUDE SMALL (LESS THAN 5%) SHIFTS IN RBC, MCV, RDW, AND MPV ALONG WITH 10 - 20 % SHIFT IN OPTICAL PLATELET COUNTS. THIS IS AN INITIAL REPORT. AN INVESTIGATION IS IN PROCESS. A FINAL REPORT WILL BE SUBMITTED WHEN THE INVESTIGATION IS COMPLETE.
THE CUSTOMER STATES THAT QC FOR THE PLATELET PARAMETER IS LOW FOR THE PLATELET OPTICAL COUNT (POC) WHEN USING A CELL-DYN 4000 ANALYZER WITH CDS/R/4K DILUENT/SHEATH REAGENT LOT #42063I2. THE CUSTOMER STATES THAT THE POC RESULTS ARE ABOUT HALF THE EXPECTED VALUE. NO SUSPECT RESULTS WERE REPORTED OUT OF THE LAB AND NO IMPACT TO PT MANAGEMENT WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CDS/R/4K DILUENT/SHEATH, 20L | DILUENT/SHEATH FOR USE ON THE CELL-DYN SAPPHIRE, CELL-DYN 40 | GKL | ABBOTT DIAGNOSTICS DIVISION/CELLDYN | NA | 42063I2 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NO INFO | CELL-DYN 4000 ANALYZER LIST NO. 1H01-01 |