FDA Adverse Event Injury Summary report: N

PRISM 6 CHANNEL ANALYZER

MDR report key: 8060205 · Received November 12, 2018

Report

Report Number
1628664-2018-02004
Event Type
Injury
Date Received
November 12, 2018
Report Date
December 19, 2018
Manufacturer
ABBOTT MANUFACTURING INC
Product Code
MZA
PMA / PMN Number
BK970042
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

FURTHER INVESTIGATION OF THE CUSTOMER ISSUE INCLUDED A REVIEW OF THE COMPLAINT TEXT, A SEARCH FOR SIMILAR COMPLAINTS, A REVIEW OF LABELING AND INSTRUMENT SERVICE. NO ADVERSE TREND WAS IDENTIFIED FOR THE CUSTOMER'S ISSUE. LABELING WAS REVIEWED AND FOUND TO BE ADEQUATE AND THE ABBOTT PRISM OPERATIONS MANUAL NOTES CAUTION AND WARNING ASSOCIATED WITH PHYSICAL HAZARDS AND EXPLAINS THE SAFETY ICONS WHICH ARE PRESENT ON THE ABBOTT PRISM SYSTEM. THE ISSUE WAS RESOLVED BY REATTACHING THE BRACKET AND BASE PLATE (ABBOTT PRISM WASTE CART TRAY, PART NUMBER 1-52433-01). THE BASE PLATE IS ATTACHED TO THE PRISM INSTRUMENT VIA A MAGNETIC BRACKET. THIS BRACKET HAD BECOME LOOSE AND FALLEN OFF CAUSING THE BASE PLATE TO MOVE WHEN THE WASTE CART WAS BEING TAKEN IN AND OUT OF THE INSTRUMENT. THE BRACKET SLOTS ONTO THE UNDERNEATH OF THE PRISM FRAME BY 4 BOLTS. THESE BOLTS HAD BECOME LOOSE AND THE BRACKET FELL OFF. BASED ON ALL AVAILABLE INFORMATION AND ABBOTT DIAGNOSTICS' COMPLAINT INVESTIGATION NO PRODUCT DEFICIENCY WAS IDENTIFIED.

Additional Manufacturer Narrative · 1

ALL AVAILABLE PATIENT INFORMATION WAS INCLUDED. ADDITIONAL PATIENT DETAILS ARE NOT AVAILABLE. AN EVALUATION IS IN PROCESS. A FINAL REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE.

Description of Event or Problem · 1

A STAFF MEMBER AT MANCHESTER BLOOD TRANSFUSION SERVICES INJURED THEIR BACK WHILE OPERATING THE PRISM ANALYZER. THE STAFF MEMBER SUFFERED A MUSCULAR BACK INJURY. THE PATIENT IS NO LONGER IN PAIN, BUT IS A LITTLE STIFF AROUND THE INJURED AREA. THE STAFF MEMBER DID NOT SEE A DOCTOR, THEY TOOK IBUPROFEN TO HELP WITH THE PAIN AND SWELLING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
899920 PRISM 6 CHANNEL ANALYZER AUTOMATED IMMUNOASSAY ANALYZER MZA ABBOTT MANUFACTURING INC

Patients

Seq Age Sex Outcome Treatment
1 Disability