ENDO CLIP III
Report
- Report Number
- 9612501-2018-01977
- Event Type
- Malfunction
- Date Received
- November 11, 2018
- Report Date
- December 27, 2018
- Manufacturer
- DAVIS & GECK CARIBE LTD
- Product Code
- FZP
- UDI-DI
- 10884521057852
- PMA / PMN Number
- K100242
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
EVALUATION SUMMARY: POST MARKET VIGILANCE (PMV) LED AN EVALUATION OF ONE DEVICE. VISUAL INSPECTION OF THE INSTRUMENT NOTED THE HANDLE WAS FLACCID AND NOT ATTACHED TO THE INTERNAL COMPONENTS. THE INSTRUMENT WAS DISMANTLED FOR VISUALIZATION OF INTERNAL COMPONENTS WHICH REVEALED THAT THE WISHBONE LINK WHICH ATTACHES THE TRIGGER TO THE FIRING MECHANISM HAD DISENGAGED. THE WISHBONE LINK WAS REATTACHED TO THE TRIGGER HANDLE AND THE INSTRUMENT WAS REASSEMBLED. THE INSTRUMENT WAS THEN FOUND TO CYCLE WITHOUT BINDING. FOURTEEN CLIPS LOADED INTO THE JAWS, FORMED PROPERLY, RELEASED FROM THE JAWS AND REMAINED SECURELY ATTACHED TO TEST MEDIA. WHEN THE CARTRIDGE WAS EMPTY, THE INTERLOCK ENGAGED TO PREVENT THE JAWS FROM APPROXIMATING. A REVIEW OF THE DEVICE HISTORY RECORD INDICATES THIS PRODUCT WAS RELEASED MEETING ALL QUALITY RELEASE SPECIFICATIONS AT THE TIME OF MANUFACTURE. REPLICATION OF THE DISENGAGED WISHBONE LINK CONDITION MAY OCCUR IF THE HANDLE IS FORCEFULLY PULLED OPEN PRIOR TO FULLY COMPLETING THE FULL HANDLE COMPRESSION. THE ROOT CAUSE OF THE OBSERVED DAMAGE WAS MISUSE OF THE PRODUCT WHICH CAUSED OR CONTRIBUTED TO THE REPORTED CONDITION. NO FURTHER ACTIONS HAVE BEEN DEEMED NECESSARY AT THIS TIME. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
ACCORDING TO THE REPORTER, DURING A MEDIASTINOSCOPY PROCEDURE, AFTER THE APPLICATION OF THE SECOND CLIP, THE CLAMP GOT STUCK IN THE CLOSER POSITION. DESPITE OF AN ATTEMPT TO RESET, THE CLAMP HAS NOT OPENED. THE DEVICE COULD BE REMOVED WITHOUT ANY OBSERVED INCIDENT. NO INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 898083 | ENDO CLIP III | CLIP, IMPLANTABLE | FZP | DAVIS & GECK CARIBE LTD | 176630 | J8E2082X | 10884521057852 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 73 YR |