FDA Adverse Event Malfunction Summary report: N

9610847-2018-00381

MDR report key: 8055190 · Received November 9, 2018

Report

Report Number
9610847-2018-00381
Event Type
Malfunction
Date Received
November 9, 2018
Date of Event
October 8, 2018
Report Date
December 12, 2018
PMA / PMN Number
N/A
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IT
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

A DEVICE HISTORY REVIEW WAS CONDUCTED FOR LOT NUMBER 8123808. OUR RECORDS SHOW THE REPORTED LOT WAS MANUFACTURED ON 05/15/2018, AND DETERMINED THAT THIS IS THE ONLY INSTANCE OF A DISCONNECTION OCCURRING IN THIS PRODUCTION BATCH. ACCORDING TO THE SAMPLING PLAN APPLIED FOR PRODUCT PERFORMANCE, THIS LOT WAS ACCEPTED AND RELEASED WITHOUT DEFECTS BEING NOTED DURING THE FINAL ASSEMBLY OR VISUAL INSPECTIONS. UNFORTUNATELY A SAMPLE COULD NOT BE OBTAINED FOR EVALUATION AND TESTING. WITHOUT THE ABILITY TO INVESTIGATE THE AFFECTED UNIT OUR QUALITY ENGINEERS WERE UNABLE TO DETERMINE THE ROOT CAUSE FOR THIS COMPLAINT. BD WAS NOT ABLE TO DUPLICATE OR CONFIRM THE CUSTOMER¿S INDICATED FAILURE MODE. CUSTOMER REPORTED DISCONNECTION AND BLOOD LEAKAGE ISSUES, HOWEVER, SAMPLE OR PHOTO WERE NOT PROVIDED WHICH ARE NECESSARY TO PERFORM A BETTER INVESTIGATION. QUALITY RECORDS HAVE BEEN CONSULTED FOR TRACKING AND TRENDING PURPOSES AND NO ISSUES LIKE THIS ARE DETECTED WHICH MEANS PRETTY LOW OCCURRENCE. PROCESS (B)(4) AND PROCESS (B)(4) WERE REVIEWED AND THERE ARE PROPER CONTROLS IN PLACE TO DETECT PRODUCT MALFUNCTIONS. BASED ON INVESTIGATION RESULTS TO DATE, ROOT CAUSE FOR MANUFACTURING PROCESS CANNOT BE DETERMINED.

Description of Event or Problem · 0

IT WAS REPORTED THAT BD CONNECTA STOPCOCK DISCONNECTED AND LEAKED. NO SERIOUS INJURY OR MEDICAL INTERVENTION WAS REPORTED.

Additional Manufacturer Narrative · 1

A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED, BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT BD CONNECTA¿ STOPCOCK DISCONNECTED AND LEAKED. NO SERIOUS INJURY OR MEDICAL INTERVENTION WAS REPORTED.

Patients

Seq Age Sex Outcome Treatment
1 Other