FDA Adverse Event Malfunction Summary report: N

INRATIO

MDR report key: 805018 · Received January 5, 2007

Report

Report Number
2954730-2006-00637
Event Type
Malfunction
Date Received
January 5, 2007
Date of Event
December 6, 2006
Report Date
January 5, 2007
Manufacturer
HEMOSENSE, INC.
Product Code
JPA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

INRATIO PRECISION DATA PROVIDED BY END-USER LOT 060570/060398: FIRST TEST INR=5.5, SECOND TEST INR = 5.9. MEAN = 5.7: SD = 0.28: %CV = 5.0%. THE %CV IS LESS THAN OR EQUAL TO 20%. PER INTERNAL PROCEDURE, THE PRECISION PASSES THE CRITERIA FOR PRECISION. THE TEST IS CONSIDERED PRECISE AND NO FURTHER TESTING IS REQUIRED AT THIS TIME. DISCREPANT RESULTS (ACCURACY) COMPARISON OF INRATIO TEST WITH LAB RESULTS PROVIDED BY END-USER AT TIME COMPLAINT WAS FILED. DATE 12/22/2006, INRATIO 5.3, LAB 4.2, MEAN 4.75, CONFIDENCE LIMITS 2.6 - 6.9. PER INTERNAL PROCEDURE, THE MEAN OF THE INRATIO METER AND COMPARATIVE SYSTEM INR WERE CALCULATED. BOTH INRATIO AND LAB VALUES ARE WITHIN THE CONFIDENCE LIMITS FOR INR TESTING. THE RESULTS ARE NOT CONSIDERED DISCREPANT WITHIN THE CONTEXT OF THE DOCUMENTED VARIABILITY FOR INR TESTING. THEREFORE FURTHER TESTING IS NOT REQUIRED AT THIS TIME.

Description of Event or Problem · 1

CALLER ALLEGED IMPRECISION WITH INRATIO. RESULTS AS FOLLOWS: FIRST TEST INR=5.5, SECOND TEST INR=5.9. CALLER ALLEGED INACCURACY WITH INRATIO. RESULTS AS FOLLOWS: DATE 12/22/2006, INRATIO 5.3, LAB 4.2, MEAN 4.75, CONFIDENCE LIMITS 2.6 - 6.9.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INRATIO PROTHROMBIN TIME TEST JPA HEMOSENSE, INC. * 060570

Patients

Seq Age Sex Outcome Treatment
1 YR