HEARTWARE VENTRICULAR ASSIST SYSTEM-PUMP
Report
- Report Number
- 3007042319-2018-05410
- Event Type
- Injury
- Date Received
- November 7, 2018
- Date of Event
- March 2, 2017
- Report Date
- January 9, 2019
- Manufacturer
- HEARTWARE, INC.
- Product Code
- DSQ
- PMA / PMN Number
- P100047
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- 003
Narratives
INVESTIGATION SUMMARY: THE PUMP AND UNKNOWN COMPONENT WITH UNKNOWN SERIAL NUMBERS WERE NOT RETURNED FOR EVALUATION. REVIEW OF LOG FILES COULD NOT BE CONDUCTED SINCE LOG FILES WERE NOT AVAILABLE FOR ANALYSIS. THERE IS NO EVIDENCE TO SUGGEST A DEVICE MALFUNCTION CAUSED OR CONTRIBUTED TO THE REPORTED EVENT. THERE IS INSUFFICIENT INFORMATION REGARDING THE REPORTED "COMPONENT MALFUNCTION" EVENT. AS A RESULT, THE REPORTED "COMPONENT MALFUNCTION" EVENT COULD NOT BE CONFIRMED. POSSIBLE CLINICAL FACTORS THAT MAY HAVE CONTRIBUTED TO THIS EVENT INCLUDE THE PATIENT¿S PRE-EXISTING HISTORY AND RELATED COMORBIDITIES, THE PROGRESSION OF THEIR UNDERLYING DISEASE, ISSUES RELATED TO THE THERAPEUTIC USE OF ANTICOAGULANT AND ANTIPLATELET MEDICATIONS AND THE PATIENT'S COMPLEX POST-OPERATIVE COURSE. THERE ARE POSSIBLE PATIENT, PHARMACOLOGICAL AND PROCEDURAL FACTORS THAT MAY HAVE CONTRIBUTED TO THIS EVENT. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
CONCOMITANT MEDICAL PRODUCTS: PRODUCT ID: UNKNOWN - MCS. OTHER DEVICES INVOLVED IN THIS EVENT: HEARTWARE VENTRICULAR ASSIST SYSTEM ¿ OTHER COMPONENT. MODEL UNKNOWN SERIAL NUMBER: UNKNOWN, EXPIRATION DATE: UNKNOWN, UDI: UNKNOWN. MFG DATE: UNKNOWN. (B)(4). IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
IT WAS REPORTED THAT THE PATIENT EXPERIENCED A PUMP THROMBOSIS AND UNSPECIFIED MALFUNCTION ON AN "OTHER COMPONENT". THE VENTRICULAR ASSIST DEVICE (VAD) AND OTHER COMPONENT REMAIN IN USE. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 889278 | HEARTWARE VENTRICULAR ASSIST SYSTEM-PUMP | VENTRICULAR (ASSISST) BYPASS | DSQ | HEARTWARE, INC. | UNK-PUMP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 56 YR | Life Threatening |