FDA Adverse Event Malfunction Summary report: N

FEMORAL TRIAL SLAPHAMMER

MDR report key: 8045970 · Received November 7, 2018

Report

Report Number
3005985723-2018-00657
Event Type
Malfunction
Date Received
November 7, 2018
Date of Event
October 10, 2018
Report Date
March 19, 2019
Manufacturer
MAKO SURGICAL CORP.
Product Code
NPJ
UDI-DI
00848486002732
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 0

REPORTED EVENT: AN EVENT REGARDING CRACK/FRACTURE INVOLVING A MAKO IMPACTOR WAS REPORTED. THE EVENT WAS CONFIRMED AFTER EXAMINATION OF THE RETURNED DEVICE. METHOD & RESULTS: PRODUCT EVALUATION AND RESULTS: VISUAL INSPECTION: VISUAL INSPECTION OF THE RETURNED DEVICE WAS PERFORMED AS PART OF THE MATERIAL ANALYSIS REPORT DATED 14 JANUARY, 2019. THE INSPECTION NOTED THE FOLLOWING: THE TIP OF THE SLAPHAMMER HAD FRACTURED, WITH THE FRACTURED TIP SCREWED INTO THE TRIAL. THE ARTICULATING SURFACE OF THE TRIAL AND FRACTURED SLAPHAMMER TIP ARE SHOWN IN FIGURE 3. THE FRACTURE SURFACES OF THE SLAPHAMMER ARE SHOWN IN FIGURES 4 AND 5. THE DIRECTION OF FRACTURE PROPAGATION COULD NOT BE DETERMINED. THE SLAPHAMMER WAS SECTIONED AS SHOWN IN FIGURE 1 FOR FURTHER ANALYSIS. THE SECTIONED SLAPHAMMER WAS ANALYZED UNDER A SCANNING ELECTRON MICROSCOPE (SEM) FOR FURTHER EVALUATION. SEM ANALYSIS: SEM IMAGES OF THE SLAPHAMMER FRACTURE SURFACE ARE SHOWN IN FIGURES 6 THROUGH 10. AN OVERVIEW OF THE FRACTURE SURFACE IS SHOWN IN FIGURE 6. TRANSGRANULAR CLEAVAGE AND DUCTILE DIMPLE FRACTURE MORPHOLOGIES WERE OBSERVED ON THE FRACTURE SURFACE, INDICATING THE DEVICE FRACTURED IN OVERLOAD (FIGURES 7 THROUGH 9). INTERGRANULAR FRACTURE MORPHOLOGIES WERE ALSO OBSERVED ON THE FRACTURE SURFACE, ALSO INDICATING A FRACTURE DUE TO OVERLOAD (FIGURE 10). ENERGY-DISPERSIVE SPECTROSCOPY (EDS) ANALYSIS WAS PERFORMED ON THE SLAPHAMMER (FIGURE 11). ELEMENTAL CONSTITUENTS OF THE SLAPHAMMER TIP INCLUDED FE-CR-NI-TI-MO, WHICH ARE CONSISTENT WITH 465 PH STAINLESS STEEL ALLOY. MATERIAL ANALYSIS: MATERIAL ANALYSIS OF THE RETURNED DEVICE NOTED THE FOLLOWING: THE SLAPHAMMER TIP HAD FRACTURED. MIXED-MODE AND INTERGRANULAR FRACTURE MORPHOLOGIES WERE OBSERVED ON THE SLAPHAMMER, CONSISTENT WITH AN OVERLOAD CONDITION. THE SLAPHAMMER MET THE DRAWING REQUIREMENT FOR HARDNESS. THE INTERGRANULAR FRACTURE MORPHOLOGIES ARE POTENTIALLY DUE TO HYDROGEN EMBRITTLEMENT. EDS SHOWED THE SLAPHAMMER WAS CONSISTENT WITH 465 PH STAINLESS STEEL ALLOY. BASED ON THE GIVEN INFORMATION, NO IDENTIFIABLE MATERIALS OR MANUFACTURING DISCREPANCIES WERE OBSERVED ON THE SURFACES EXAMINED. MEDICAL RECORDS RECEIVED AND EVALUATION: NOT PERFORMED AS NO MEDICAL INFORMATION WAS RECEIVED FOR REVIEW. PRODUCT HISTORY REVIEW: A REVIEW OF THE DEVICE MANUFACTURING RECORDS INDICATE THAT ALL DEVICES ACCEPTED INTO FINAL STOCK WERE FREE FROM DISCREPANCIES. -COMPLAINT HISTORY REVIEW: THERE HAVE BEEN 4 OTHER SIMILAR EVENTS FOR THE LOT REFERENCED. CONCLUSIONS: THE EXACT CAUSE OF THE EVENT COULD NOT BE DETERMINED BECAUSE INSUFFICIENT INFORMATION WAS PROVIDED. FURTHER INFORMATION SUCH AS THE PRIMARY OPERATIVE REPORT IS NEEDED TO COMPLETE THE INVESTIGATION FOR DETERMINING ROOT CAUSE. NO FURTHER INVESTIGATION FOR THIS EVENT IS POSSIBLE AT THIS TIME. IF ADDITIONAL INFORMATION BECOMES AVAILABLE TO INDICATE FURTHER EVALUATION IS WARRANTED, THIS RECORD WILL BE REOPENED.

Description of Event or Problem · 0

FEMORAL TRIAL HAD A PIECE OF THE FEMORAL IMPACTOR THAT BROKE OFF DURING IMPACTION. CASE TYPE: PKA.

Additional Manufacturer Narrative · 1

REVIEW OF THE DEVICE HISTORY RECORDS INDICATE DEVICES WERE MANUFACTURED AND ACCEPTED INTO FINAL STOCK WITH NO RELEVANT REPORTED DISCREPANCIES. BASED ON THE DEVICE IDENTIFICATION THE COMPLAINT DATABASES WERE REVIEWED FOR SIMILAR REPORTED EVENTS. THERE HAVE BEEN THREE OTHER EVENTS FOR THE REPORTED LOT WHICH ALSO RELATE TO THE CRACK/FRACTURE OF A SLAP HAMMER. A SUPPLEMENTAL REPORT WILL BE SUBMITTED UPON COMPLETION OF THE INVESTIGATION.

Description of Event or Problem · 1

FEMORAL TRIAL HAD A PIECE OF THE FEMORAL IMPACTOR THAT BROKE OFF DURING IMPACTION. CASE TYPE: PKA.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
887580 FEMORAL TRIAL SLAPHAMMER PROSTHESIS, KNEE PATELLOFEMOROTIBIAL, PARTIAL, SEMI-CONSTRAINED, CEMENTED, POLYM NPJ MAKO SURGICAL CORP. 06060515 00848486002732

Patients

Seq Age Sex Outcome Treatment
1 Other