FDA Adverse Event Injury Summary report: N

LOTUS INTRODUCER SET

MDR report key: 8042517 · Received November 6, 2018

Report

Report Number
3004193842-2018-00010
Event Type
Injury
Date Received
November 6, 2018
Date of Event
October 3, 2018
Report Date
November 6, 2018
Manufacturer
CREGANNA MEDICAL ALSO D.B.A CREGANNA TACTX MEDICAL
Product Code
DYB
PMA / PMN Number
K140338
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE COMPLAINT PRODUCT WAS NOT RETURNED TO CREGANNA MEDICAL AND IT WAS THEREFORE NOT POSSIBLE TO CARRY OUT ANY DIMENSIONAL, FUNCTIONAL, VISUAL OR MICROSCOPIC EXAMINATION ON THE DEVICE. AS SUCH, THE REPORTED VESSEL DISSECTION CANNOT BE CONFIRMED. A REVIEW OF RISK MANAGEMENT DOCUMENTATION WAS COMPLETED. BASED ON THIS REVIEW AND INFORMATION AVAILABLE, IT WAS CONCLUDED THAT THERE IS NO INDICATION OF A POTENTIAL PROCESSING OR DESIGN FAILURE ASSOCIATED WITH THIS COMPLAINT AND THEREFORE NO UPDATES ARE REQUIRED TO THE RISK DOCUMENTATION DUE TO THE REPORTED INCIDENT. A REVIEW OF THE MANUFACTURING DOCUMENTATION FOR LOT# 389778 AND ALL OF ITS SUBCOMPONENTS WAS COMPLETED AND FOUND THAT THE DEVICE MET ITS MATERIAL, ASSEMBLY AND PRODUCT SPECIFICATIONS AT THE TIME OF RELEASE TO DISTRIBUTION. THE REVIEW OF THE ROUTER AND SUBSEQUENT SUB ASSEMBLY ROUTERS DID NOT HIGHLIGHT ANY ANOMALIES. THERE WERE NO CAPAS (CORRECTIVE AND PREVENTATIVE ACTION), MRRS (MATERIAL REVIEW REPORT) OR DEVIATIONS GENERATED FOR THIS LOT, #389778 DURING THE MANUFACTURING PROCESS THAT MAY HAVE CONTRIBUTED TO THE REPORTED ISSUE. AS OF 06TH NOVEMBER 2018 WHEN THE REVIEW WAS COMPLETED, THERE WAS ONE OTHER COMPLAINT ASSOCIATED WITH LOT NUMBER 389778, LOT-PC18-022 TW# (B)(4) FOR THE AS REPORTED FAILURE LOTUS - PATIENT - VESSEL DISSECTION. THIS HAS BEEN FILED UNDER MDR REPORT# 3004193842-2018-00011. FROM THE INFORMATION AVAILABLE, THERE IS NO EVIDENCE PRESENT TO INDICATE THAT THE DEVICE WAS NOT USED PER THE DIRECTIONS FOR USE/PRODUCT LABEL. FOLLOWING COMPLETION OF CREGANNA MEDICAL INFORMATION REVIEW, THE COMPLAINT ANALYSED CLASSIFICATION HAS BEEN ASSIGNED AS:" LOTUS - PRODUCT NOT RETURNED - COMPLAINT UNABLE TO CONFIRM". BASED ON THE COMPLAINT REVIEW AND THE EVENT DESCRIPTION FOR THIS COMPLAINT, THE ROOT CAUSE CLASSIFICATION ASSIGNED TO THIS COMPLAINT IS 'ANTICIPATED PROCEDURAL COMPLICATION' WHICH IS DEFINED AS 'THE COMPLAINT IS DUE TO A KNOWN PHYSIOLOGICAL EFFECT OF THE PROCEDURE NOTED WITHIN THE INSTRUCTIONS FOR USE, AND OR DEVICE LABELLING' VESSEL DISSECTION IS ANTICIPATED PROCEDURAL COMPLICATION AND ARE KNOWN PHYSIOLOGICAL EFFECT OF THE PROCEDURE AS NOTED WITHIN THE INSTRUCTIONS FOR USE. THIS COMPLAINT WAS ESCALATED TO THE QUALITY MANAGEMENT TEAM. BASED ON THE ABOVE CONCLUSION NO FURTHER ESCALATION OR CORRECTIVE ACTION IS REQUIRED AT THIS TIME. WE WILL CONTINUE TO MONITOR FOR THESE COMPLAINT TYPES.

Description of Event or Problem · 1

COMPLAINT DESCRIPTION RECEIVED AT CREGANNA MEDICAL IS AS FOLLOWS: "IT WAS REPORTED THAT SIMILARLY WHEN THEY TOOK THE LOTUS INTRODUCER OUT OF THE PATIENT THE ARTERY HAD A DISSECTION AND WAS FIXED BY UTILIZING TWO 8X5MM VIABOND STENTS . PATIENT IS FINE. NO COMPLICATIONS."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
880613 LOTUS INTRODUCER SET INTRODUCER CATHETER DYB CREGANNA MEDICAL ALSO D.B.A CREGANNA TACTX MEDICAL H749NTR200 389778

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention