FDA Adverse Event Malfunction Summary report: N

LMA PROSEAL, REU, SIZE 3 (150030)

MDR report key: 8041466 · Received November 6, 2018

Report

Report Number
9681900-2018-00039
Event Type
Malfunction
Date Received
November 6, 2018
Date of Event
October 8, 2018
Report Date
October 30, 2018
Manufacturer
TELEFLEX MEDICAL
Product Code
CAE
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SN
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

(B)(4). LOT# CORRECTED TO SN: (B)(4), CATALOG# CORRECTED TO 15140. THE SAMPLE WAS RETURNED FOR EVALUATION. A VISUAL EXAM WAS PERFORMED AND IT WAS OBSERVED THAT THERE WERE BLUE STAINS AND A PINK MARK ON THE CUFF. IT WAS ALSO NOTICED THAT THE AIRWAY TUBE WAS YELLOWISH. UPON CLOSER EXAMINATION OF THE FAILURE LOCATION IT WAS FOUND THERE WERE JAGGED EDGES ON THE CONNECTOR. THE JAGGED EDGES AT WHERE THE CONNECTOR SPLIT APPEARED TO BE RUPTURED BY FORCE. A DEVICE HISTORY RECORD REVIEW WAS PERFORMED AND THERE WERE NO ISSUES FOUND THAT COULD RELATE TO THE REPORTED COMPLAINT. BASED ON THE INVESTIGATION PERFORMED, THE REPORTED COMPLAINT WAS CONFIRMED. THE USE OF NON-COMPATIBLE DETERGENT OR CONCENTRATED DETERGENT ON THE LMA DEVICE COULD HAVE POSSIBLY ADVERSELY IMPACTED THE STURDINESS OF THE CONNECTOR, WHICH COULD RESULT IN THE CONNECTOR BREAKING WHEN EXTERNAL FORCE IS APPLIED.

Description of Event or Problem · 0

CUSTOMER COMPLAINT ALLEGES "THE PROSEAL 15MM CONNECTOR BROKE OFF AFTER CONNECTING IT TO THEIR VENTILATOR". NO PATIENT INJURY OR CONSEQUENCE REPORTED. PATIENT CONDITION REPORTED AS FINE.

Additional Manufacturer Narrative · 1

QN# (B)(4).

Description of Event or Problem · 1

CUSTOMER COMPLAINT ALLEGES "THE PROSEAL 15MM CONNECTOR BROKE OFF AFTER CONNECTING IT TO THEIR VENTILATOR". NO PATIENT INJURY OR CONSEQUENCE REPORTED. PATIENT CONDITION REPORTED AS FINE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
881974 LMA PROSEAL, REU, SIZE 3 (150030) AIRWAY,OROPHARYNGEAL,ANESTHESIOLOGY CAE TELEFLEX MEDICAL 8WSAK9S4

Patients

Seq Age Sex Outcome Treatment
1 VENTILATOR| VENTILATOR