FDA Adverse Event Malfunction Summary report: N

HEATER-COOLER SYSTEM, 3T

MDR report key: 8040833 · Received November 6, 2018

Report

Report Number
9611109-2018-01316
Event Type
Malfunction
Date Received
November 6, 2018
Date of Event
October 8, 2018
Report Date
November 6, 2018
Manufacturer
LIVANOVA DEUTSCHLAND
Product Code
DWC
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THERE WAS NO PATIENT INVOLVEMENT. THE HEATER-COOLER 16-02-80 IS NOT DISTRIBUTED IN THE USA, BUT IT IS SIMILAR TO HEATER-COOLER 16-02-85, WHICH IS DISTRIBUTED IN THE USA (510(K) NUMBER: K052601). (B)(4) MANUFACTURES THE HEATER-COOLER SYSTEM 3T. THE INCIDENT OCCURRED IN (B)(6). IF ANY ADDITIONAL INFORMATION PERTINENT TO THE REPORTED EVENT IS RECEIVED, IT WILL BE PROVIDED IN A SUPPLEMENTAL REPORT. DEVICE NOT RETURNED.

Description of Event or Problem · 1

(B)(4) RECEIVED A REPORT THAT A HEATER-COOLER SYSTEM 3T DISPLAYED AN ERROR MESSAGE ON PATIENT SIDE DURING MAINTENANCE. THERE WAS NO PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
884264 HEATER-COOLER SYSTEM, 3T CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS DWC LIVANOVA DEUTSCHLAND 16-02-80

Patients

Seq Age Sex Outcome Treatment
1