CATALYS SYSTEM
Report
- Report Number
- 3005675890-2018-00102
- Event Type
- Malfunction
- Date Received
- November 2, 2018
- Date of Event
- October 4, 2018
- Report Date
- November 4, 2020
- Manufacturer
- JOHNSON & JOHNSON SURGICAL VISION, INC
- Product Code
- OOE
- UDI-DI
- 05050474609792
- PMA / PMN Number
- K121091
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KY, US
- Reporter Occupation
- PHYSICIAN
Narratives
THIS CORRECTION MDR IS BEING SUBMITTED TO ADDRESS INACCURATE INFORMATION IN SECTION F ON THE XML VERSION OF THE PREVIOUS SUBMISSION CAUSED BY A SOFTWARE GLITCH IN THE TRANSMISSION PROCESS. THE INTERNAL NON-CONFORMANCE NUMBERS FOR THIS SOFTWARE ISSUE ARE NR-0148919 AND CAPA: 010215.
JOHNSON & JOHNSON SURGICAL VISION, INC INVESTIGATION SUBSEQUENTLY IDENTIFIED THAT THE CATALYS SYSTEM MAY HAVE A REMOTE POTENTIAL EVENT WHERE LOSS OF SUCTION DURING LASER FIRING COULD CREATE A HAZARDOUS SITUATION THAT MAY RESULT IN SCORING OF THE POSTERIOR CORNEAL SURFACE. THEREFORE JOHNSON & JOHNSON SURGICAL VISION, INC HAS ISSUED AN ADVISORY NOTICE TO REINFORCE INSTRUCTIONS PROVIDED IN PRODUCT TRAINING AND IN THE OPERATOR¿S MANUAL REGARDING SUCTION LOSS DURING A PROCEDURE. A REVIEW OF RECORDS RELATED TO THE DEVICE INCLUDING LABELING, COMPLAINT TRENDING, AND RISK DOCUMENTATION WILL BE PERFORMED. UPON COMPLETION OF THIS REVIEW, IF THERE IS ANY FURTHER RELEVANT INFORMATION OBTAINED, A SUPPLEMENTAL MEDWATCH WILL BE FILED. ALL PERTINENT INFORMATION AVAILABLE TO JOHNSON & JOHNSON SURGICAL VISION, INC HAS BEEN SUBMITTED.
IT WAS REPORTED THAT VACUUM WAS LOST DURING PRIMARY INCISION TREATMENT IN LEFT EYE. PRIMARY INCISION ABORTED AND SIDE PORT INCISION NOT DONE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 873902 | CATALYS SYSTEM | CATALYS | OOE | JOHNSON & JOHNSON SURGICAL VISION, INC | CATALYS-U | 05050474609792 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 72 YR |