FDA Adverse Event Injury Summary report: N

10MM/130 DEG TI CANN TFNA 340MM/RIGHT - STERILE

MDR report key: 8033744 · Received November 2, 2018

Report

Report Number
2939274-2018-54748
Event Type
Injury
Date Received
November 2, 2018
Report Date
October 17, 2018
Manufacturer
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
Product Code
HSB
UDI-DI
10886982096104
PMA / PMN Number
K160167
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. MANUFACTURING LOCATION: MONUMENT; MANUFACTURING DATE: SEPTEMBER 09, 2016; EXPIRATION DATE: AUGUST 31, 2026; PART: 04.037.054S; 10MM/130 DEG TI CANN TFNA 340MM/RIGHT ¿ STERILE; LOT: H182957 (STERILE); LOT QUANTITY: (B)(4) WORK ORDER TRAVELER MET ALL INSPECTION ACCEPTANCE CRITERIA. INSPECTION SHEET, IN-PROCESS / INSPECT DIMENSIONAL / FINAL MET ALL INSPECTION ACCEPTANCE CRITERIA. INSPECTION SHEET, TFNA ASSEMBLY INSPECTION MET ALL INSPECTION ACCEPTANCE CRITERIA. PACKAGING LABEL LOG LPPF WAS REVIEWED AND DETERMINED TO BE CONFORMING. PACKAGING BOM WAS REVIEWED AND FOUND TO BE CONFORMING WITH NO DEVIATIONS TO NORMAL PACKAGING IDENTIFIED. STERILIZATION CONTROL NUMBER (SCN) 12949 SUPPLIED BY ETHICON (ABQ) WAS REVIEWED AND DETERMINED TO BE CONFORMING. THIS LOT MET ALL DIMENSIONAL, VISUAL, STERILIZATION AND PACKAGING CRITERIA AT THE TIME OF RELEASE WITH NO ISSUES DOCUMENTED DURING THE MANUFACTURE THAT WOULD CONTRIBUTE TO THIS COMPLAINT CONDITION. COMPONENT PARTS REVIEWED: PART: 04.037.942.2; LOCK PRONG 130 DEGREE, TFNA, BP55; LOT: L037875; LOT QUANTITY: (B)(4) PURCHASED FINISHED GOODS TRAVELER MET ALL INSPECTION ACCEPTANCE CRITERIA. PART: 04.037.912.4; WAVE SPRING, SHIM ENDED, BP55; LOT: H045932; LOT QUANTITY: (B)(4) WORK ORDER TRAVELER MET ALL INSPECTION ACCEPTANCE CRITERIA. INSPECTION SHEET, INCOMING FINAL INSPECTION MET ALL INSPECTION ACCEPTANCE CRITERIA. MATERIAL CERTIFICATION AND CERTIFICATE OF CONFORMANCE AND QUALITY HISTORY CARD RECEIVED FROM SMALLEY WERE REVIEWED AND DETERMINED TO BE CONFORMING. PART: 04.037.912.3; TFNA LOCK DRIVE, BP58; LOT: H161535; LOT QUANTITY: (B)(4) WORK ORDER TRAVELER MET ALL INSPECTION ACCEPTANCE CRITERIA. INSPECTION SHEET MET ALL INSPECTION ACCEPTANCE CRITERIA. PART: 21127; TIMOAGRI16.00, BP80; LOT: H086530; LOT QUANTITY: (B)(4). CERTIFIED TEST REPORT RECEIVED FROM PERRYMAN COMPANY AND CERTIFICATE OF ANALYSIS RECEIVED FROM METALWERKS WERE REVIEWED AND DETERMINED TO BE CONFORMING. LOT SUMMARY REPORT MET ALL INSPECTION ACCEPTANCE CRITERIA. RAW MATERIAL RECEIVING/PUTAWAY CHECKLIST MET ALL INSPECTION ACCEPTANCE CRITERIA. PRODUCT WAS NOT RETURNED. BASED ON THE INFORMATION AVAILABLE, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND PREVENTATIVE ACTION IS PROPOSED. THIS COMPLAINT WILL BE ACCOUNTED FOR AND MONITORED VIA POST MARKET SURVEILLANCE ACTIVITIES. IF ADDITIONAL INFORMATION IS MADE AVAILABLE, THE INVESTIGATION WILL BE UPDATED AS APPLICABLE. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 0

DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE. COMPLAINANT PART IS NOT EXPECTED TO BE RETURNED FOR MANUFACTURER REVIEW/INVESTIGATION PER HOSPITAL CONTACT REPORTER. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS.

Additional Manufacturer Narrative · 0

DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. UPDATED FOR INVESTIGATION BASED ON PROVIDED X-RAY. DEVICE HISTORY LOT MANUFACTURING LOCATION: MONUMENT, MANUFACTURING DATE: 09-SEP-2016, EXPIRATION DATE: 31-AUG-2026, PART NUMBER: 04.037.054S, 10MM/130 DEG TI CANN TFNA 340MM/RIGHT ¿ STERILE, LOT NUMBER: H182957 (STERILE). THIS LOT MET ALL DIMENSIONAL, VISUAL, STERILIZATION AND PACKAGING CRITERIA AT THE TIME OF RELEASE WITH NO ISSUES DOCUMENTED DURING THE MANUFACTURE THAT WOULD CONTRIBUTE TO THIS COMPLAINT CONDITION. VISUAL INSPECTION: DEVICE WAS NOT RETURNED. VISUAL INSPECTION OF THE PROVIDED X-RAY IMAGE CONFIRMED THE CONDITION OF POST OPERATIVE NAIL BREAKAGE, WHICH AGREES WITH THE REPORTED COMPLAINT CONDITION. THE X-RAY SHOWS THAT THE NAIL HAS BROKEN AT THE PROXIMAL HOLE. THE BREAK APPEARS OBLIQUE. NO NEW MALFUNCTIONS WERE IDENTIFIED AS A RESULT OF THIS INVESTIGATION. REVIEW OF THE DEVICE HISTORY RECORD(S) SHOWED THAT THERE WERE NO ISSUES DURING THE MANUFACTURE OF THE PRODUCT THAT WOULD CONTRIBUTE TO THIS COMPLAINT CONDITION. DOCUMENT/SPECIFICATION REVIEW: NO PRODUCT DESIGN ISSUES OR DISCREPANCIES WERE OBSERVED DURING THIS INVESTIGATION. CONCLUSION: A DEFINITIVE ROOT CAUSE FOR THE NAIL BREAKING POSTOPERATIVELY COULD NOT BE DETERMINED BASED ON THE PROVIDED INFORMATION. THIS COMPLAINT IS CONFIRMED HOWEVER NO PRODUCT DESIGN ISSUES OR MANUFACTURING DISCREPANCIES THAT WOULD CONTRIBUTE TO THE REPORTED COMPLAINT CONDITION WERE IDENTIFIED DURING THIS INVESTIGATION AND THE DCRM ADEQUATELY ADDRESSES THE COMPLAINT CONDITION. NO NEW, UNIQUE OR DIFFERENT PATIENT HARMS WERE IDENTIFIED AS A RESULT OF THIS EVALUATION. BASED ON THE INVESTIGATION FINDINGS, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 1

IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE. EXPIRATION DATE IS UNKNOWN. COMPLAINANT PART IS EXPECTED TO BE RETURNED FOR MANUFACTURER REVIEW/INVESTIGATION, BUT HAS YET TO BE RECEIVED. (B)(4). THE INVESTIGATION COULD NOT BE COMPLETED; NO CONCLUSION COULD BE DRAWN, AS NO PRODUCT WAS RECEIVED A REVIEW OF THE DEVICE HISTORY RECORDS HAS BEEN REQUESTED. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS.

Description of Event or Problem · 1

IT WAS REPORTED THAT ON AN UNKNOWN DATE, A PATIENT UNDERWENT A REMOVAL OF A BROKEN TROCHANTERIC FEMORAL NAIL ADVANCE (TFNA) NAIL. IT WAS UNKNOWN IF THERE WAS SURGICAL DELAY. PROCEDURE AND PATIENT OUTCOME IS UNKNOWN. CONCOMITANT DEVICE REPORTED: UNKNOWN HELICAL BLADE (PART # UNKNOWN, LOT # UNKNOWN, QUANTITY UNKNOWN). UNKNOWN LOCKING SCREW (PART # UNKNOWN, LOT # UNKNOWN, QUANTITY UNKNOWN). THIS COMPLAINT INVOLVES ONE (1) DEVICE. THIS REPORT IS 1 OF 1 FOR (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
873791 10MM/130 DEG TI CANN TFNA 340MM/RIGHT - STERILE ROD,FIXATION,INTRAMEDULLARY HSB WRIGHTS LANE SYNTHES USA PRODUCTS LLC 04.037.054S H182957 10886982096104

Patients

Seq Age Sex Outcome Treatment
1 77 YR Required Intervention SEE EVENT DESCRIPTION