TRIAGE CARDIAC TRIPLE MARKER PANEL
Report
- Report Number
- 3013982035-2018-00030
- Event Type
- Malfunction
- Date Received
- November 1, 2018
- Date of Event
- September 30, 2018
- Report Date
- November 1, 2018
- Manufacturer
- QUIDEL CARDIOVASCULAR INC.
- Product Code
- MMI
- PMA / PMN Number
- K030286
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CO
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
INVESTIGATION CONCLUSION: THE CUSTOMER'S COMPLAINT WAS NOT REPLICATED DURING IN-HOUSE TESTING WITH RETAINS OF DEVICE LOT W64250RB. NO ISSUES WITH TROPONINI RECOVERY WERE OBSERVED; LOT PERFORMED PROPERLY. MANUFACTURING BATCH RECORDS FOR THE LOT WERE REVIEWED, LOT MET RELEASE SPECIFICATIONS. BASED ON THE INFORMATION AVAILABLE, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY AND NO CORRECTIVE ACTION IS REQUIRED. ALTHOUGH THIS CATALOG NUMBER (97000HSEU) IS NOT APPROVED IN THE UNITED STATES, THIS EVENT IS BEING REPORTED AS THE DEVICE IS SAME/SIMILAR TO CATALOG NUMBER 97000HS, 510(K) NUMBER K030286.
THE EVENT OCCURRED IN (B)(6). THE CUSTOMER REPORTED THE FOLLOWING OCCURRING FOR A PATIENT. ON (B)(6) AT 8AM THE DOCTOR REQUESTED A POINT OF CARE TRIAGE CARDIAC PANEL BE PERFORMED ON A PATIENT. THE PATIENT SAMPLE WAS PROCESSED AND RESULTED A TRIAGE TNI OF <0.05NG/ML AT 8:33AM, REPORTED AS "NORMAL". AT 9:38AM A "CRITICAL" RESULT OF A QUANTITATIVE TROPONIN WITH A VALUE OF 0.2160NG/ML WAS RECEIVED. THE SAME SAMPLE WAS REPROCESSED ON BOTH PLATFORMS. ON TRIAGE AT 11:26AM THE SAMPLE YIELDED A TNI OF <0.05NG/ML. ON THE QUANTITATIVE TROPONIN AT 2:34PM A TNI OF 0.2010NG/ML WAS RECEIVED. TRIAGE CUT-OFF IS 0.4NG/ML. CUT-OFF FOR QUANTITATIVE TNI NOT REPORTED, ONLY STATED RESULTS WERE CRITICAL. PATIENT UNDERWENT CARDIAC CATHETERIZATION. DESPITE MULTIPLE ATTEMPTS, FURTHER INFORMATION IS NOT AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 870818 | TRIAGE CARDIAC TRIPLE MARKER PANEL | TRIAGE CARDIAC TRIPLE MARKER PANEL | MMI | QUIDEL CARDIOVASCULAR INC. | 97000HSEU | W64250RB |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 81 YR |