FDA Adverse Event Malfunction Summary report: N

PNEUPAC PARAPAC VENTILATOR

MDR report key: 8030884 · Received November 1, 2018

Report

Report Number
3012307300-2018-08038
Event Type
Malfunction
Date Received
November 1, 2018
Date of Event
October 1, 2018
Report Date
October 8, 2019
Manufacturer
SMITHS MEDICAL ASD, INC.
Product Code
BTL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

EVALUATION RESULTS: ONE PARAPAC PLUS (PERT NUMBER 530A1166NJP/ SERIAL NUMBER (B)(4)) WAS RECEIVED WITH A BATTERY AND PATIENT CIRCUIT (LOT 160718). NO DAMAGE WAS NOTED TO THE UNIT OR THE CIRCUIT DURING VISUAL INSPECTION. FUNCTIONAL TESTING FOUND NO FAULT WITH THE DEVICE OR PATIENT CIRCUIT. ALL FUNCTIONS WERE FOUND TO BE CORRECT AND WITHIN SPECIFICATIONS. THE INVESTIGATOR TRIED TO RECREATE THE FOLLOWING REPORTED PRODUCT PROBLEM: "DURING MAINTENANCE OF THE PRODUCT, THE CUSTOMER NOTICED THE AIR SUPPLY FOR INHALATION BREATH WOULD NOT STOP AND COULD NOT BE SWITCHED TO THAT FOR EXHALATION BREATH EVEN WHEN THE MODE WAS SWITCHED TO "AIR VENTILATION".' THE UNIT WAS REPEATEDLY TESTED AT VARIOUS INPUT PRESSURES RANGING FROM 41-90 PSI (INPUT GAS PRESSURE RANGE STATED ON THE UNIT). THE UNIT CONTINUED TO FUNCTION CORRECTLY AT ALL SETTINGS. THE REPORTED FAULT COULD NOT BE CONFIRMED; THE UNIT FUNCTIONED CORRECTLY AND WITHIN SPECIFICATIONS AT ALL SETTINGS.

Additional Manufacturer Narrative · 1

REPORT SOURCE: (B)(6).

Description of Event or Problem · 1

INFORMATION WAS RECEIVED THAT A SMITHS MEDICAL PNEUPAC PARAPAC VENTILATOR FAILED TO STOP THE AIR INTAKE SUPPLY DURING MAINTENANCE TESTING. ADDITIONALLY, THE VENTILATOR WOULD NOT SWITCH TO "EXPIRATION," EVEN WHEN THE MODE WAS SWITCHED TO "AIR VENTILATION". NO ADVERSE PATIENT EFFECTS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
869579 PNEUPAC PARAPAC VENTILATOR VENTILATOR, EMERGENCY, POWERED (RESUSCITATOR) BTL SMITHS MEDICAL ASD, INC.

Patients

Seq Age Sex Outcome Treatment
1