PALMAZ GENESIS 6 X 12, 135CM
Doing adverse event research by hand?
The Research Assistant searches MAUDE for you, groups reports by failure mode, and cites every record.
Free to try · every answer cites its records
Report
- Report Number
- 9610978-2007-00007
- Event Type
- Injury
- Date Received
- January 5, 2007
- Date of Event
- July 27, 2006
- Report Date
- July 27, 2006
- Manufacturer
- CORDIS EUROPA, N.V.
- Product Code
- FGE
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- RISK MANAGER
Narratives
THE PRODUCT WAS NOT RETURNED TO CORDIS FOR ANALYSIS. A DEVICE HISTORY RECORD REVIEW WAS PERFORMED AND SHOWED THAT THIS LOT OF PRODUCTS MET ALL REQUIREMENTS PER THE APPLICABLE MANUFACTURING QUALITY PLAN (MQP). THIS REVIEW WAS EXTENDED TO SUBASSEMBLY E7136899 WITH LOT NUMBER 0311040562 AND TO SUBASSEMBLY E7144773 WITH LOT NUMBERS 0104042758 AND 0111040247. THIS REVIEW CONFIRMED THAT SUBASSEMBLIES MET ALL REQUIREMENTS PER THE APPLICABLE MQP AS WELL, INCLUDING BURST TEST VALUES. WITH THE LIMITED AMOUNT OF INFO AVAILABLE AND WITHOUT THE RETURN OF THE PRODUCT OR FILMS OF THE PROCEDURE, IT IS NOT POSSIBLE TO DETERMINE THE RELATIONSHIP BETWEEN THE DEVICE AND THE EVENT. HOWEVER, IN THIS CASE, IT APPEARS THAT VESSEL CHARACTERISTICS MAY HAVE CONTRIBUTED TO THE REPORTED EVENT.
IT WAS REPORTED THAT THE BALLOON BURST. THE INTENDED PROCEDURE WAS STENTING OF THE LEFT RENAL ARTERY. THE VESSEL HAD MILD CALCIFICATION, SEVERE TORTUOSITY AND A 90% STENOSIS. THE LESION WAS WIRED WITHOUT DIFFICULTY. THE PALMAZ GENESIS STENT DELIVERY SYSTEM (SDS) WAS ADVANCED TO THE LESION AND POSITIONED. THE BALLOON WAS INFLATED WITH AN INDEFLATOR AND STENT DEPLOYMENT WAS INITIATED; HOWEVER, THE BALLOON RUPTURED AT UNK ATMOSPHERES. ANOTHER BALLOON CATHETER WAS USED TO POST-DILATED THE STENT. NO ADDITIONAL LESION/VESSEL OR PROCEDURAL DETAILS WERE PROVIDED. IT IS UNK IF ANY DIFFICULTY WAS EXPERIENCED WHILE ADVANCING THE SDS TO THE LESION OR WHILE CROSSING THE LESION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PALMAZ GENESIS 6 X 12, 135CM | * | FGE | CORDIS EUROPA, N.V. | NA | R1104180 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | * | Required Intervention | GUIDEWIRE (RADIFOCUS, 0.035/TERUMO) |