SPINNING SPIROS CLOSED MALE LUER, RED CAP
Report
- Report Number
- 9617594-2018-00152
- Event Type
- Malfunction
- Date Received
- November 1, 2018
- Date of Event
- October 18, 2018
- Report Date
- October 18, 2018
- Manufacturer
- ICU MEDICAL DE MEXICO, S. DE R.L. DE C.V.
- Product Code
- FPA
- UDI-DI
- 00840619026615
- PMA / PMN Number
- K082806
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
THE DEVICE WILL NOT BE RETURNING FOR EVALUATION. THE REPORTED COMPLAINT CANNOT BE CONFIRMED WITHOUT THE RETURNED SAMPLE. A DEVICE HISTORY REVIEW (DHR) WAS COMPLETED AND NO NON-CONFORMANCES WERE NOTED. CONCOMITANT MEDICAL PRODUCT: MAXPLUS NEEDLELESS CONNECTOR, MFR BECTON, DICKINSON AND COMPANY; THERAPY DATE: (B)(6) 2018. 45" EXTENSION SET WITH A/S VALVE, LN AMS 435CV-1, LOT NUM: 1806004D, MFR: VYGON; RAPIDFILL CONNECTOR, LN H93813901, LOT #: 60123404, MFR BAXTER; 600CC SYRINGE, MFR BECTON, DICKINSON AND COMPANY, THERAPY DATE: UNKNOWN.
THE EVENT INVOLVED A SPINNING SPIROS CLOSED MALE LUER, RED CAP THAT A LEAK OF FLUROURACIL-5FU WAS NOTICED ON AN AMBULATORY PUMP SETUP THAT WAS IN USE FOR 46 HOURS. THE SPIROS WAS ATTACHED TO A MAXPLUS NEEDLELESS VALVE. IT WAS REPORTED THERE WAS NOTICEABLE WHITE RESIDUE POWDER IN/ON THE SPIROS AND OUTSIDE OF THE MAXPLUS-SPIROS CONNECTION. THE CUSTOMER REPORTED THAT THE CASSETTE RESERVOIR WAS PREPARED IN THE PHARMACY USING A VYGON EXTENSION SET TO THE CASSETTE AND PRIMED MANUALLY WITH A BAXTER RAPIDFILL CONNECTOR AND A BECTON, DICKINSON 600CC SYRINGE; POSSIBLE MISALIGNED CONNECTION WITH RAPIDFILL CONNECTOR AND SPIROS BEING FULLY ENGAGED. THERE WAS PATIENT INVOLVEMENT, NO ADVERSE EVENTS, NO MEDICAL OR SURGICAL INTERVENTIONS PROVIDED AND NO DELAY IN CRITICAL THERAPY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 867895 | SPINNING SPIROS CLOSED MALE LUER, RED CAP | SPINNING SPIROS | FPA | ICU MEDICAL DE MEXICO, S. DE R.L. DE C.V. | CH2000S-C | 3761680 | 00840619026615 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |