FDA Adverse Event Malfunction Summary report: N

SPINNING SPIROS CLOSED MALE LUER, RED CAP

MDR report key: 8027480 · Received November 1, 2018

Report

Report Number
9617594-2018-00152
Event Type
Malfunction
Date Received
November 1, 2018
Date of Event
October 18, 2018
Report Date
October 18, 2018
Manufacturer
ICU MEDICAL DE MEXICO, S. DE R.L. DE C.V.
Product Code
FPA
UDI-DI
00840619026615
PMA / PMN Number
K082806
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WILL NOT BE RETURNING FOR EVALUATION. THE REPORTED COMPLAINT CANNOT BE CONFIRMED WITHOUT THE RETURNED SAMPLE. A DEVICE HISTORY REVIEW (DHR) WAS COMPLETED AND NO NON-CONFORMANCES WERE NOTED. CONCOMITANT MEDICAL PRODUCT: MAXPLUS NEEDLELESS CONNECTOR, MFR BECTON, DICKINSON AND COMPANY; THERAPY DATE: (B)(6) 2018. 45" EXTENSION SET WITH A/S VALVE, LN AMS 435CV-1, LOT NUM: 1806004D, MFR: VYGON; RAPIDFILL CONNECTOR, LN H93813901, LOT #: 60123404, MFR BAXTER; 600CC SYRINGE, MFR BECTON, DICKINSON AND COMPANY, THERAPY DATE: UNKNOWN.

Description of Event or Problem · 1

THE EVENT INVOLVED A SPINNING SPIROS CLOSED MALE LUER, RED CAP THAT A LEAK OF FLUROURACIL-5FU WAS NOTICED ON AN AMBULATORY PUMP SETUP THAT WAS IN USE FOR 46 HOURS. THE SPIROS WAS ATTACHED TO A MAXPLUS NEEDLELESS VALVE. IT WAS REPORTED THERE WAS NOTICEABLE WHITE RESIDUE POWDER IN/ON THE SPIROS AND OUTSIDE OF THE MAXPLUS-SPIROS CONNECTION. THE CUSTOMER REPORTED THAT THE CASSETTE RESERVOIR WAS PREPARED IN THE PHARMACY USING A VYGON EXTENSION SET TO THE CASSETTE AND PRIMED MANUALLY WITH A BAXTER RAPIDFILL CONNECTOR AND A BECTON, DICKINSON 600CC SYRINGE; POSSIBLE MISALIGNED CONNECTION WITH RAPIDFILL CONNECTOR AND SPIROS BEING FULLY ENGAGED. THERE WAS PATIENT INVOLVEMENT, NO ADVERSE EVENTS, NO MEDICAL OR SURGICAL INTERVENTIONS PROVIDED AND NO DELAY IN CRITICAL THERAPY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
867895 SPINNING SPIROS CLOSED MALE LUER, RED CAP SPINNING SPIROS FPA ICU MEDICAL DE MEXICO, S. DE R.L. DE C.V. CH2000S-C 3761680 00840619026615

Patients

Seq Age Sex Outcome Treatment
1