FDA Adverse Event Other Summary report: N

SHAMPAINE

MDR report key: 802513 · Received December 1, 2004

Report

Report Number
3019090-2004-00002
Event Type
Other
Date Received
December 1, 2004
Report Date
November 4, 2004
Manufacturer
GETINGE USA, INC, F.K.A MDT DIAGNOSTIC E.
Product Code
FQO
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PILOT OPERATED CHECK VALVE WAS REPLACED AND PROPER OPERATION OF THE TABLE WAS VERIFIED. REPRESENTATIVE EXAMINED THE VALVE AND THE CONCLUSION IS THAT THE DRIFTING WAS MOST LIKELY CAUSED BY A PIECE OF CONTAMINANT BECOMING LODGED BETWEEN THE SEAL AND THE PISTON AT THE INSTANT THAT THE PISTON CLOSED ALLOWING FLUID TO FLOW.

Description of Event or Problem · 1

DURING SURGERY, THE TABLE WAS PUT INTO SLIGHT TRENDELENBURG POSITION BY OR STAFF. AFTER MOTOR WAS OFF THE TABLE DRIFTED INTO FULL TRENDELENBURG POSITION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SHAMPAINE SURGICAL TABLE FQO GETINGE USA, INC, F.K.A MDT DIAGNOSTIC E. 5100B *

Patients

Seq Age Sex Outcome Treatment
1 58 YR