466P306X
Report
- Report Number
- 1016427-2018-02179
- Event Type
- Malfunction
- Date Received
- October 31, 2018
- Date of Event
- August 14, 2018
- Report Date
- May 30, 2019
- Manufacturer
- CORDIS CASHEL
- Product Code
- DTK
- PMA / PMN Number
- K020316
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
CORRECTION TO SECTION COMMON DEVICE NAME (PRODUCT CODE).
THIS ARTICLE WAS FOUND DURING A RECENT LITERATURE SEARCH OF THIS DEVICE. PLEASE NOTE THAT PATIENT SPECIFIC DETAILS (MEDICAL HISTORY AND REASON FOR INTERVENTION) ARE NOT AVAILABLE. THE DEVICE IS A TRAPEASE FILTER AND THE CATALOG AND LOT NUMBERS ARE NOT AVAILABLE. KOIZUMI, J., HARA, T., SEKIGUCHI, T., ICHIKAWA, T., TAJIMA, H., TAKENOSHITA, N., SHIRATO, K. (2018). MULTICENTER INVESTIGATION OF THE INCIDENCE OF INFERIOR VENA CAVA FILTER FRACTURE. JAPANESE JOURNAL OF RADIOLOGY, 36(11), 661-668. DOI:10.1007/S11604-018-0764-X AS NOTED IN THE PUBLICATION BY KOIZUMI ET AL., MULTICENTER INVESTIGATION OF THE INCIDENCE OF INFERIOR VENA CAVA FILTER FRACTURE; JPN J RADIOL. 2018 AUG 14. DOI: 10.1007/S11604-018-0764-X; REPORT ONE CASE OF TRAPEASE FILTER FRACTURE 856 DAYS POST IMPLANTATION. THE FRACTURED STRUTS ARE LOCATED IN THE UPPER AND LOWER UMBRELLAS. THE FRAGMENT MIGRATED TO THE PULMONARY ARTERY. THE FILTER WAS IMPLANTED IN THE SUPRARENAL POSITION. THE PRODUCT WAS NOT RETURNED FOR ANALYSIS AND THE STERILE LOT NUMBER HAS NOT BEEN PROVIDED; THEREFORE, NO DEVICE ANALYSIS NOR DEVICE HISTORY RECORD REVIEW COULD BE PERFORMED. WITHOUT PROCEDURAL FILMS AVAILABLE FOR REVIEW, THE REPORTED ¿FILTER- FRACTURED-SEPARATED - IN PATIENT¿ AND ¿FILTER- MIGRATION¿ COULD NOT BE CONFIRMED AND THE EXACT CAUSE COULD NOT BE DETERMINED. THE INSTRUCTIONS FOR USE (IFU) STATES FILTER FRACTURE IS A POTENTIAL COMPLICATION OF VENA CAVA FILTERS. ANATOMIC LOCATIONS THAT CREATE CONCENTRATED STRESS POINTS FROM FILTER DEFORMATION (FOR EXAMPLE, DEPLOYMENT AT APEX OF SCOLIOSIS, OVERLAPPING OF EITHER OF THE RENAL OSTIA, OR PLACEMENT ADJACENT TO A VERTEBRAL OSTEOPHYTE) MAY CONTRIBUTE TO FRACTURE OF A PARTICULAR FILTER STRUT. INFERIOR VENA CAVA (IVC) FILTER MIGRATION IS A KNOWN POTENTIAL ADVERSE EVENT ASSOCIATED WITH ALL IVC FILTER IMPLANTS AND IS LISTED IN THE INSTRUCTION FOR USE (IFU) AS SUCH. POSSIBLE CAUSES FOR FILTER MIGRATION INCLUDES MEGA CAVA, WIRE ENTRAPMENT DURING CENTRAL VENOUS CATHETER PLACEMENT, ¿SAIL¿ EFFECT (CRANIAL MIGRATION) OF LARGE CLOT BURDEN WITHIN THE FILTER, MECHANICAL DEVICE FAILURE, AND OPERATOR ERROR. PHYSIOLOGIC CAUSES OF MIGRATION MAY RESULT FROM TEMPORARY DYSMORPHISM OF THE INFERIOR VENA CAVA INCLUDING BENDING, COUGHING OR VALSALVA MANEUVERS RESULTING IN DISLODGMENT OF THE FILTER. SOME STUDIES SUGGEST THAT STRENUOUS PHYSICAL ACTIVITY AND INCREASED INTRA-ABDOMINAL PRESSURE CAN LEAD TO MIGRATION OF IVC FILTERS. GIVING THE LIMITED INFORMATION AVAILABLE AT THIS TIME, CLINICAL FACTOR CONTRIBUTING TO THE MIGRATION COULD NOT BE DETERMINED. ALSO, WITH THE LIMITED INFORMATION AVAILABLE FOR REVIEW AT THIS TIME, THERE IS NOTHING TO SUGGEST THAT THE REPORTED EVENT IS RELATED TO THE DESIGN AND/OR MANUFACTURING PROCESS OF THE DEVICE; THEREFORE NO CORRECTIVE ACTION WILL BE TAKEN. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THE FILE WILL BE UPDATED ACCORDINGLY.
AS NOTED IN THE PUBLICATION BY KOIZUMI ET AL., MULTICENTER INVESTIGATION OF THE INCIDENCE OF INFERIOR VENA CAVA FILTER FRACTURE; JPN J RADIOL. 2018 AUG 14. DOI: 10.1007/S11604-018-0764-X; REPORT ONE CASE OF TRAPEASE FILTER FRACTURE 856 DAYS POST IMPLANTATION. THE FRACTURED STRUTS ARE LOCATED IN THE UPPER AND LOWER UMBRELLAS. THE FRAGMENT MIGRATED TO THE PULMONARY ARTERY. THE FILTER WAS IMPLANTED IN THE SUPRARENAL POSITION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 866536 | 466P306X | FILTER, INTRAVASCULAR, CARDIOVASCULAR | DTK | CORDIS CASHEL | UNKNOWN |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 51 YR |