Narratives
BATCH REVIEW PERFORMED ON 31 OCTOBER 2018: LOT 178447: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 03-APRIL-2018. EXPIRATION DATE: 2023-03-23. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT. ADDITIONAL COMPONENTS INVOLVED (DISLOCATION) AND REVISED: LINER: CC E CC LIGHT 01.26.3239 HCT FLAT PE HC LINER Ø 32 / C, (K120531). LOT 144764: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 22-SEPTEMBER-2014. EXPIRATION DATE: 2019-08-31. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT. BALL HEADS: MECTACER 01.29.206 BIOLOX DELTA CERAMIC BALL HEAD 12/14 Ø 32 SIZE L +4, (K112115). LOT 162726: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 27-JULY-2016. EXPIRATION DATE: 2021-07-19. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT. CLINICAL EVALUATION PERFORMED BY MEDICAL AFFAIRS DIRECTOR: FEW DAYS AFTER PRIMARY CEMENTLESS TOTAL HIP ARTHROPLASTY, THE STEM ROTATED IN THE FEMUR AND THE HIP DISPLACED. THE STEM COULDN'T FIND INITIAL FIXATION IN THE BONE THAT HAS A DORR C CONFIGURATION AND A LOW DENSITY. WE DO NOT SEE ANY REASON WHY THIS LACK OF INITIAL STABILITY COULD BE ASCRIBED TO A DEFECT IN THE DEVICE.
PATIENT CAME IN COMPLAINING OF PAIN 7 DAYS AFTER PRIMARY. THE STEM ROTATED IN THE FEMORAL CANAL DUE TO POOR BONE QUALITY. THE HEAD DISLOCATED FROM THE LINER. THE SURGEON REVISED ALL MEDACTA COMPONENTS AND REVISED WITH A MEDACTA BIPOLAR SYSTEM. THE SURGERY WAS PERFORMED SUCCESSFULLY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 866271 | STEM: AMISTEM H HA COATED STD STEM SIZE 2 | HIP CEMENTLESS STEM | LZO | MEDACTA INTERNATIONAL SA | 178447 | 07630030804083 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |