FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-523RNAB

MDR report key: 8022549 · Received October 31, 2018

Report

Report Number
2032227-2018-52791
Event Type
Malfunction
Date Received
October 31, 2018
Date of Event
October 14, 2018
Report Date
January 21, 2019
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
UDI-DI
00643169513051
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MS, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

INSULIN PUMP RECEIVED WITH INTERMITTENT BUTTON RESPONSE DUE TO PARTIAL UNLOCKED KEYPAD CONNECTOR NOTED DURING VISUALLY INSPECT. UNABLE TO PERFORM ALL FUNCTIONAL TESTING INCLUDING THE DISPLACEMENT, OPERATING CURRENTS, REWIND, BASIC OCCLUSION, OCCLUSION, PRIME AND EXCESSIVE NO DELIVERY TEST DUE TO BUTTONS NOT RESPONDING.

Additional Manufacturer Narrative · 1

(B)(4). CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

THE CUSTOMER REPORTED VIA PHONE CALL THAT THEY EXPERIENCED HIGH BLOOD GLUCOSE LEVEL AND THE INSULIN PUMP HAD A BUTTON ERROR AND BUTTONS WERE NOT WORKING. THE CUSTOMER¿S BLOOD GLUCOSE LEVEL WAS 402, 223, 180 AND 360 MG/DL. ADDITIONALLY THE BLOOD GLUCOSE VALUE WAS 180MG/DL WHICH SHOT UP TO 300MG/DL. CUSTOMER DECLINED TO TROUBLESHOOTING FOR HIGH BLOOD GLUCOSE LEVEL. CUSTOMER WAS ADVISED TO OBSERVE TIME CLOCK FOR ONE MINUTE TO VERIFY IF TIME ADVANCES AND THEY STATED THAT THE TIME WAS ADVANCES. THE CUSTOMER WAS ADVISED TO DISCONTINUE USE OF THE INSULIN PUMP AND TO REVERT TO THE BACK-UP PLAN. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
865408 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-523RNAB PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-523RNAB A1523RNABJ 00643169513051

Patients

Seq Age Sex Outcome Treatment
1 49 YR