FDA Adverse Event Malfunction Summary report: N

SERVO-U

MDR report key: 8020577 · Received October 30, 2018

Doing adverse event research by hand?

The Research Assistant searches MAUDE for you, groups reports by failure mode, and cites every record.

Free to try · every answer cites its records

Report

Report Number
8010042-2018-00576
Event Type
Malfunction
Date Received
October 30, 2018
Report Date
January 9, 2019
Manufacturer
MAQUET CRITICAL CARE AB
Product Code
CBK
PMA / PMN Number
K151814
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

GETINGE USA SALES, LLC (IMPORTER) IS SUBMITTING THIS REPORT ON BEHALF OF MAQUET CRITICAL CARE AB (MANUFACTURER). REF. EXEMPTION #: E2018003 GETINGE USA SALES, LLC 45 BARBOUR POND DRIVE WAYNE, NJ 07470 CONTACT PERSON: (B)(4).

Additional Manufacturer Narrative · 0

GETINGE USA SALES, LLC (IMPORTER) IS SUBMITTING THIS REPORT ON BEHALF OF MAQUET CRITICAL CARE AB (MANUFACTURER). REF. EXEMPTION #: E2018003. GETINGE USA SALES, LLC (B)(4). CONTACT PERSON:(B)(4). THE VENTILATOR WAS INVESTIGATED ON SITE AND THE O2 GAS MODULE WAS REPLACED AND RETURNED FOR MANUFACTURE INVESTIGATION. THE REPORTED DRIFTING OF O2 CONCENTRATION WERE REPRODUCED DURING TEST OF THE RETURNED O2 GAS MODULE MOUNTED IN A VENTILATOR. THE DEVIATION WAS CAUSED BY A NOZZLE UNIT IN THE O2 GAS MODULE. IF OSCILLATIONS HAPPEN DURING VENTILATION, THE PATIENT MAY RECEIVE AN AMOUNT OF OXYGEN IN THE GAS MIXTURE THAT DEVIATE FROM THE EXPECTED. FLOW AND PRESSURE MAY ALSO DEVIATE FROM THE EXPECTED. ALARMS WILL BE ACTIVATED IF THE FAILURE OCCURS DURING VENTILATION. EVALUATION OF THE RECEIVED DEVICE LOGS SHOWS THAT DURING THE EVENT THE VENTILATION MODE WAS NEURALLY ADJUSTED VENTILATORY ASSIST(NAVA). THERE ARE FOUR MATCHING EVENTS ON OCTOBER 6, THERE THE FIRST ALARM FOR O2 CONCENTRATION HIGH APPEARING 11:31 AND THE LAST ALARM 22:25.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING VENTILATION, THE O2 CONCENTRATION WAS DRIFTING. THERE WAS NO PATIENT HARM. (B)(4).

Description of Event or Problem · 0

IMPORTER REF. #: (B)(4). MANUFACTURER REF. #: (B)(4).