FDA Adverse Event Injury Summary report: N

QUADRA-H CEMENTLESS, HA COATED LAT STEM SIZE 2

MDR report key: 8019251 · Received October 30, 2018

Report

Report Number
3005180920-2018-00850
Event Type
Injury
Date Received
October 30, 2018
Date of Event
October 9, 2018
Report Date
October 31, 2018
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
JDI
UDI-DI
07630030802140
PMA / PMN Number
K082792
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
SZ
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

ON 30 OCTOBER 2018 WE WERE INFORMED THAT THE SUSPECTED PE WEAR IS NOT CONFIRMED. THE EVENT DESCRIPTION IS REVISION SURGERY PERFORMED, 9 YEARS AFTER PRIMARY SURGERY, DUE TO ASEPTIC FEMORAL LOOSENING.

Additional Manufacturer Narrative · 1

BATCH REVIEW PERFORMED ON 30 OCTOBER 2018. LOT 081458: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 10 JULY 2008. EXPIRATION DATE: 2013-05-31. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITHOUT ANY SIMILAR REPORTED EVENT. OTHER DEVICE INVOLVED: LINER: CC E CC LIGHT FLAT PE LINER Ø 28 / E REFERENCE 01.26.2844STT. LOT 060220: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 18 APRIL 2006 . EXPIRATION DATE: 2011-02-28. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT. CLINICAL EVALUATION PERFORMED BY MEDICA AFFARIS MANAGER: HIP REVISION SURGERY OCCURRED 9 YEARS AFTER PRIMARY CEMENTLESS TOTAL HIP ARTHROPLASTY. RADIOGRAPHIC IMAGES PROVIDED SHOW THE PRESENCE OF RADIOLUCENT LINES AROUND THE STEM AND SIGNS OF STRESS SHIELDING. ASEPTIC LOOSENING IS A POSSIBLE, LITERATURE DESCRIBED ADVERSE EVENT AFTER CEMENTLESS TOTAL HIP ARTHROPLASTY AND CAUSES ARE OFTEN UNKNOWN. THE REASON OF THIS EVENT CANNOT BE DETERMINED.

Description of Event or Problem · 1

ON (B)(6) 2018 WE WERE INFORMED THAT A REVISION SURGERY WILL BE PERFORMED DUE TO ASEPTIC FEMORAL LOOSENING, THE REASON IS PROBABLY PE WEAR.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
860893 QUADRA-H CEMENTLESS, HA COATED LAT STEM SIZE 2 CEMENTLESS HIP STEM JDI MEDACTA INTERNATIONAL SA 081458 07630030802140

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention