QUADRA-H CEMENTLESS, HA COATED LAT STEM SIZE 2
Report
- Report Number
- 3005180920-2018-00850
- Event Type
- Injury
- Date Received
- October 30, 2018
- Date of Event
- October 9, 2018
- Report Date
- October 31, 2018
- Manufacturer
- MEDACTA INTERNATIONAL SA
- Product Code
- JDI
- UDI-DI
- 07630030802140
- PMA / PMN Number
- K082792
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SZ
- Reporter Occupation
- 003
Narratives
ON 30 OCTOBER 2018 WE WERE INFORMED THAT THE SUSPECTED PE WEAR IS NOT CONFIRMED. THE EVENT DESCRIPTION IS REVISION SURGERY PERFORMED, 9 YEARS AFTER PRIMARY SURGERY, DUE TO ASEPTIC FEMORAL LOOSENING.
BATCH REVIEW PERFORMED ON 30 OCTOBER 2018. LOT 081458: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 10 JULY 2008. EXPIRATION DATE: 2013-05-31. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITHOUT ANY SIMILAR REPORTED EVENT. OTHER DEVICE INVOLVED: LINER: CC E CC LIGHT FLAT PE LINER Ø 28 / E REFERENCE 01.26.2844STT. LOT 060220: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 18 APRIL 2006 . EXPIRATION DATE: 2011-02-28. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT. CLINICAL EVALUATION PERFORMED BY MEDICA AFFARIS MANAGER: HIP REVISION SURGERY OCCURRED 9 YEARS AFTER PRIMARY CEMENTLESS TOTAL HIP ARTHROPLASTY. RADIOGRAPHIC IMAGES PROVIDED SHOW THE PRESENCE OF RADIOLUCENT LINES AROUND THE STEM AND SIGNS OF STRESS SHIELDING. ASEPTIC LOOSENING IS A POSSIBLE, LITERATURE DESCRIBED ADVERSE EVENT AFTER CEMENTLESS TOTAL HIP ARTHROPLASTY AND CAUSES ARE OFTEN UNKNOWN. THE REASON OF THIS EVENT CANNOT BE DETERMINED.
ON (B)(6) 2018 WE WERE INFORMED THAT A REVISION SURGERY WILL BE PERFORMED DUE TO ASEPTIC FEMORAL LOOSENING, THE REASON IS PROBABLY PE WEAR.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 860893 | QUADRA-H CEMENTLESS, HA COATED LAT STEM SIZE 2 | CEMENTLESS HIP STEM | JDI | MEDACTA INTERNATIONAL SA | 081458 | 07630030802140 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |