FREESTYLE LIBRE
Report
- Report Number
- 2954323-2018-08065
- Event Type
- Injury
- Date Received
- October 25, 2018
- Date of Event
- September 20, 2018
- Report Date
- May 8, 2019
- Manufacturer
- ABBOTT DIABETES CARE INC.
- Product Code
- PZE
- PMA / PMN Number
- P160030
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- 003
Narratives
NO PRODUCT HAS BEEN RETURNED. EXTENDED INVESTIGATION HAS BEEN PERFORMED FOR THE REPORTED COMPLAINT AND THERE WAS NO INDICATION THAT THE PRODUCT DID NOT MEET SPECIFICATION. THE REPORTED COMPLAINT IS RELATED TO SKIN IRRITATION OR AN ALLERGIC REACTION TO THE PATCH ADHESIVE OF THE FREESTYLE LIBRE SENSOR. DOSE AUDIT AND ENVIRONMENTAL MONITORING REPORTS WERE REVIEWED FOR ISSUES RELATING TO STERILITY OF THE PRODUCT. THE SENSOR COMPONENT HAS NO IMPACT ON PRODUCT STERILITY AND THEREFORE, SENSOR COMPONENTS DHRS WERE NOT REVIEWED. SENSOR KIT DHRS HAVE BEEN REVIEWED TO ASSESS THE MANUFACTURING PROCESS, WHICH INCLUDES THE APPLICATION OF THE ADHESIVE TO THE PUCK. A DHR (DEVICE HISTORY REVIEW) FOR THE FREESTYLE LIBRE SENSOR KIT WAS REVIEWED AND THE DHR SHOWED THE FREESTYLE LIBRE SENSOR KIT PASSED ALL TESTS PRIOR TO RELEASE. DOSE AUDIT REPORTS WERE REVIEWED AND DEMONSTRATES THE CONTINUED EFFECTIVENESS OF THE ESTABLISHED STERILIZATION PROCESS FOR LIBRE SENSOR KITS. ENVIRONMENTAL MONITORING REPORTS WERE REVIEWED, INCLUDING BIOBURDEN AND ENDOTOXIN TESTING, AND DEMONSTRATED THAT ALL MONITORING PROCESSES CONTINUE TO MEET ADC MINIMUM REQUIREMENTS FOR PRODUCT QUALITY. ALL REVIEW ACTIVITIES CONDUCTED ABOVE, INCLUDING BUT NOT LIMITED TO THE FINAL RELEASE TESTING SPECIFICALLY ASSOCIATED WITH THE MANUFACTURE OF THIS PRODUCT, ARE SUFFICIENT INFORMATION IN ORDER TO SHOW IF THE PRODUCT HAS MET SPECIFICATIONS PRIOR TO RELEASE. IF THE PRODUCT IS RETURNED, THE CASE WILL BE RE-OPENED AND A PHYSICAL INVESTIGATION WILL BE PERFORMED.
CUSTOMER REPORTED EXPERIENCING AN ADVERSE SKIN REACTION AFTER 8 DAYS OF WEARING AN ADC FREESTYLE LIBRE SENSOR, AND REPORTED FIRST OBSERVING REDNESS ON (B)(6) 2018. CUSTOMER HAD CONTACT WITH A HEALTHCARE PROFESSIONAL ON (B)(6) 2018 AND WAS PRESCRIBED FENISTIL (ANTIHISTAMINE) WITH CORTISONE FOR TREATMENT. THERE WAS NO REPORT OF DEATH OR PERMANENT INJURY ASSOCIATED WITH THIS EVENT.
THE REPORTED PRODUCT IS NOT EXPECTED TO BE RETURNED AS THE CUSTOMER DISCARDED THE SENSOR. EXTENDED INVESTIGATION IS PENDING. A FOLLOW-UP REPORT WILL BE FILED IF PRODUCT IS RETURNED OR ADDITIONAL INFORMATION IS OBTAINED. THE DEVICE MFG DATE IS UNKNOWN. THE DATE ENTERED IS THE DATE ABBOTT DIABETES CARE BECAME AWARE OF THE EVENT. ALL PERTINENT INFORMATION AVAILABLE TO ABBOTT DIABETES CARE HAS BEEN SUBMITTED.
CUSTOMER REPORTED EXPERIENCING AN ADVERSE SKIN REACTION AFTER 8 DAYS OF WEARING AN ADC FREESTYLE LIBRE SENSOR, AND REPORTED FIRST OBSERVING REDNESS ON (B)(6) 2018. CUSTOMER HAD CONTACT WITH A HEALTHCARE PROFESSIONAL ON (B)(6) 2018 AND WAS PRESCRIBED FENISTIL (ANTIHISTAMINE) WITH CORTISONE FOR TREATMENT. THERE WAS NO REPORT OF DEATH OR PERMANENT INJURY ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 844555 | FREESTYLE LIBRE | FLASH GLUCOSE MONITORING SYSTEM | PZE | ABBOTT DIABETES CARE INC. | 71702-01 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |