TRUEMETRIX
Report
- Report Number
- 1000113657-2018-01100
- Event Type
- Malfunction
- Date Received
- October 25, 2018
- Date of Event
- October 2, 2018
- Report Date
- October 25, 2018
- Manufacturer
- TRIVIDIA HEALTH, INC.
- Product Code
- NBW
- UDI-DI
- 00021292007690
- PMA / PMN Number
- K140100
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN
- Reporter Occupation
- OTHER
Narratives
(B)(4). PRODUCT NOT YET RETURNED FOR EVALUATION. MOST LIKELY UNDERLYING ROOT CAUSE: MLC-58-USER HAD AN INACCURATE REFERENCE: SELF: THE PERSON IS USING THEMSELVES AS THE REFERENCE OR HOW THEY FEEL AT THE TIME THEY RUN THE BLOOD TEST. TEST STRIP UDI# (B)(4). NOTE: MANUFACTURER CONTACTED CUSTOMER (SEVERAL ATTEMPTS) IN A FOLLOW-UP CALL TO ENSURE THAT THE REPLACEMENT PRODUCTS RESOLVED THE INITIAL CONCERN - UNABLE TO ESTABLISH CONTACT WITH THE CUSTOMER AT THIS TIME. PRODUCT NOTIFICATION LETTER SENT TO CUSTOMER TO CONTACT CUSTOMER CARE.
CONSUMER REPORTED COMPLAINT FOR HIGH AND ERRATIC BLOOD GLUCOSE TEST RESULTS. THE CUSTOMER IS CONCERNED WITH TESTS RESULTS FROM RESULTS OBTAINED OF 178, 251, 142, 365 AND 233MG/DL; RESULTS OF 251 AND 142 MG/DL WERE BACK TO BACK TESTS. THE CUSTOMER'S EXPECTED FASTING BLOOD GLUCOSE TEST RESULT RANGE IS 100 - 140 MG/DL. THE CUSTOMER FEELS WELL AND DID NOT REPORT ANY SYMPTOMS. MEDICAL ATTENTION IS NOT REPORTED AS A RESULT OF THE ACTUAL BLOOD GLUCOSE RESULTS. DURING THE CALL ON (B)(6) 2018, A BACK TO BACK BLOOD TEST WAS PERFORMED BY THE CUSTOMER FASTING AND PRODUCED TEST RESULTS OF 279 MG/DL AND 278 MG/DL USING TRUEMETRIX METER. THE PRODUCT IS STORED ACCORDING TO SPECIFICATION IN THE DINING ROOM. THE TEST STRIP LOT MANUFACTURER'S EXPIRATION DATE IS 03/15/2020 AND OPEN VIAL DATE IS (B)(6) 2018. THE METER MEMORY WAS REVIEWED FOR PREVIOUS TEST RESULT HISTORY: (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 846898 | TRUEMETRIX | BLOOD GLUCOSE SYSTEM | NBW | TRIVIDIA HEALTH, INC. | TRUEMETRIX | MV2928 | 00021292007690 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 0 YR |