DEVICE #1 ADVANCED CONTROL BASE, JACKSON MODULAR TABLE SYST
Report
- Report Number
- 2921578-2006-00005
- Event Type
- Other
- Date Received
- December 22, 2006
- Date of Event
- October 20, 2006
- Report Date
- December 22, 2006
- Manufacturer
- ORTHOPEDIC SYSTEMS, INC.
- Product Code
- BWN
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- INVALID DATA
Narratives
DEVICE WAS REPAIRED. USER ALLOWED MAINTENANCE OF TABLE TO GO BEYOND SAFE USE. REPAIR ORDER CREATED, ITEMS FOR REPAIR IDENTIFIED REPLACEMENT ACTUATOR AND KNOB REPLACEMENT ASSEMBLY. TABLE IS CURRENTLY IN USE.
THE PT WAS UNDERGOING HIP ARTHROPLASTY, AND AFTER THE DISTRACTION OF THE RIGHT HIP JOINT WAS APPLIED, AN ATTEMPT WAS MADE TO TILT THE TABLE FOR BETTER ACCESS. THE ANESTHESIOLOGITS WAS UNFAMILIAR WITH THE BED AND DID NOT KNOW IT WOULD NOT ROTATE 180 DEGREES. SHE WAS UNABLE TO GET THE BED TO TILT USING THE HAND CONTROL AND SEEING THE "LOCK/UNLOCK" LEVER ON THE BED, SHE UNLOCKED IT THINKING THAT WAS WHY THE BED WOULD NOT TILT. THE TABLETOP ROTATED AND THE PT WAS LEFT DANGLING BY THE TWO SAFETY STRAPS (ACROSS CHEST AND UNAFFECTED LEG). THE HEALTHCARE TEAM REACTED SWIFTYLY AND WAS ABLE TO MAINTAIN THE SECURITY OF HIS AIRWAY AND REORIENT THE BED. HE WAS MOVED TO ANOTHER BED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | DEVICE #1 ADVANCED CONTROL BASE, JACKSON MODULAR TABLE SYST | OPERATING ROOM TABLE | BWN | ORTHOPEDIC SYSTEMS, INC. | 5803 | * |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 37 YR | Other |