FDA Adverse Event Malfunction Summary report: N

ALARIS® SYRINGE MODULE SET

MDR report key: 8001554 · Received October 24, 2018

Report

Report Number
9616066-2018-02037
Event Type
Malfunction
Date Received
October 24, 2018
Date of Event
October 2, 2018
Report Date
October 3, 2018
Manufacturer
CAREFUSION
Product Code
FPA
UDI-DI
10885403233920
PMA / PMN Number
K811885
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

THE CUSTOMER'S REPORT OF A LEAK WAS CONFIRMED. FUNCTIONAL TESTING SHOWED THAT FLUID WAS LEAKING FROM THE ENGAGEMENT BETWEEN THE OUTLET PORT OF THE FILTER AND THE MICROBORE TUBING COMPONENT. FURTHER INSPECTION SHOWED THAT THE TUBING WAS NOT FULLY BONDED TO ITS MATED FILTER COMPONENT. THE ROOT CAUSE OF THE LEAK WAS A MANUFACTURING ASSEMBLY ISSUE OF INSUFFICIENT SOLVENT APPLICATION AT THE ENGAGEMENT.

Description of Event or Problem · 0

THE CUSTOMER REPORTED DURING A FENTANYL AND DOPAMINE INFUSION, A LEAK WAS NOTED BELOW BOTH 0.2 MICRON FILTERS ON THE EXTENSION SETS. IT WAS REPORTED BY THE CLINICIAN THAT BOTH SETS WERE NEW WHEN THE LEAK OCCURRED. THERE WAS NO REPORT OF PATIENT HARM. THE EVENT OCCURRED IN PICU.

Additional Manufacturer Narrative · 1

CONCOMITANT MEDICAL PRODUCTS: 8015; 10014916; (2)MZ1000-07; (2)SPM TUBING; (2)8110; (2) NON BD 12ML SYRINGE WITH A 3ML LABEL, LOT 180930-010, EXP (B)(6) 2018, 0.45% SODIUM CHLORIDE FLUSH; THERAPY DATE (B)(6) 2018. THE AFFECTED PRODUCT HAS BEEN RECEIVED AND THE EVALUATION IS PENDING. A FOLLOW UP REPORT WILL BE SUBMITTED ONCE THE EVALUATION IS COMPLETED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED DURING A FENTANYL AND DOPAMINE INFUSION, A LEAK WAS NOTED BELOW BOTH 0.2 MICRON FILTERS ON THE EXTENSION SETS. IT WAS REPORTED BY THE CLINICIAN THAT BOTH SETS WERE NEW WHEN THE LEAK OCCURRED. THERE WAS NO REPORT OF PATIENT HARM. THE EVENT OCCURRED IN PICU.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
838874 ALARIS® SYRINGE MODULE SET SET, ADMINISTRATION, INTRAVASCULAR FPA CAREFUSION 10014916 18057691 10885403233920

Patients

Seq Age Sex Outcome Treatment
1 5 YR