FDA Adverse Event
Malfunction
Summary report: N
FREESTYLE FLASH
MDR report key: 799856
·
Received December 20, 2006
Report
- Report Number
- 2954323-2006-01672
- Event Type
- Malfunction
- Date Received
- December 20, 2006
- Date of Event
- November 1, 2006
- Report Date
- December 20, 2006
- Manufacturer
- ABBOTT DIABETES CARE INC.
- Product Code
- LFR
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SW
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
COMPLAINT CONFIRMED. TESTED RETURNED UNIT. NO MFG PARAMETERS FOUND OUT OF SPEC. DID NOT OBSERVE BATTERY ICON OR BOOKLET ICON WHILE STRIP WAS INSERTED. HOWEVER, UOM WAS SELECTABLE AND WAS RECEIVED AT MMOL/LS. ERRORS 015, 0204, 0300, 0800 AND 0806 WERE OBSERVED IN THE ERROR LOG INDICATING BATTERY DROOP. METER IS OPERABLE. CUSTOMERS AND RETAILERS HAVE BEEN INFORMED THROUGH THE FA16MAY2006 LETTER.
Description of Event or Problem · 1
A CUSTOMER REPORTED THAT THE UOM SETTING OF THEIR LOCKED FREESTYLE METER CHANGED FROM MMOL/L TO MG/DL. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY OR MISTREATMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | FREESTYLE FLASH | BLOOD GLUCOSE MONITORING SYSTEM | LFR | ABBOTT DIABETES CARE INC. | * | * |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | YR |