FDA Adverse Event Malfunction Summary report: N

FREESTYLE FLASH

MDR report key: 799856 · Received December 20, 2006

Report

Report Number
2954323-2006-01672
Event Type
Malfunction
Date Received
December 20, 2006
Date of Event
November 1, 2006
Report Date
December 20, 2006
Manufacturer
ABBOTT DIABETES CARE INC.
Product Code
LFR
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SW
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

COMPLAINT CONFIRMED. TESTED RETURNED UNIT. NO MFG PARAMETERS FOUND OUT OF SPEC. DID NOT OBSERVE BATTERY ICON OR BOOKLET ICON WHILE STRIP WAS INSERTED. HOWEVER, UOM WAS SELECTABLE AND WAS RECEIVED AT MMOL/LS. ERRORS 015, 0204, 0300, 0800 AND 0806 WERE OBSERVED IN THE ERROR LOG INDICATING BATTERY DROOP. METER IS OPERABLE. CUSTOMERS AND RETAILERS HAVE BEEN INFORMED THROUGH THE FA16MAY2006 LETTER.

Description of Event or Problem · 1

A CUSTOMER REPORTED THAT THE UOM SETTING OF THEIR LOCKED FREESTYLE METER CHANGED FROM MMOL/L TO MG/DL. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY OR MISTREATMENT.