MENTOR SILTEX ROUND MODERATE PROFILE
Report
- Report Number
- 1645337-2018-06348
- Event Type
- Injury
- Date Received
- October 23, 2018
- Date of Event
- October 4, 2018
- Report Date
- October 5, 2018
- Manufacturer
- MENTOR TEXAS
- Product Code
- FWM
- UDI-DI
- 00081317001768
- PMA / PMN Number
- P990075
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VA, US
- Reporter Occupation
- OTHER
Narratives
UPON RECEIPT BY MENTOR, THE DEVICE RETURNED WITHOUT FLUID. YELLOW MATERIAL WAS OBSERVED WITHIN THE DEVICE. NO FOREIGN MATERIAL WAS OBSERVED ON THE SHELL SURFACE. MENTOR PRODUCT ANALYSIS LAB DISCOVERED A RENT MEASURING APPROXIMATELY 0.3 CM WITHIN AN AREA OF SILTEX CRACKING ON THE POSTERIOR ASPECT. NO OTHER ANOMALIES WERE DISCOVERED. THE COMPLAINT WAS CONFIRMED SINCE A RUPTURE WAS FOUND ON THE DEVICE. HOWEVER, AT THIS POINT IT IS NOT POSSIBLE TO DETERMINE THE ROOT CAUSE OF SUCH DAMAGE OR THE ETIOLOGY OF THE YELLOW MATERIAL FOUND ON THE DEVICE. THE DEVICE HISTORY RECORD (DHR) FOR THE LOT NUMBER 6861915 WAS REVIEWED AND NO ANOMALIES WERE FOUND RELATED TO THIS COMPLAINT. IN ADDITION, THE DHR REVIEW VERIFIES THAT THE DEVICE WAS MANUFACTURED IN ACCORDANCE WITH DOCUMENTED SPECIFICATION AND PROCEDURES. THERE IS NO EVIDENCE THAT THE ISSUE IS RELATED WITH MANUFACTURING. SILTEX CRACKING IS DEFINED AS A MATERIAL SEPARATION OCCURRING IN THE SILTEX LAYER AND CAN EVENTUALLY PROGRESS THROUGH THE SHELL OF SILTEX DEVICES. MENTOR BELIEVES THAT SILTEX CRACKING IS ULTIMATELY A RESULT OF HIGH STRESSES CAUSED BY ACUTE FOLDS (I.E., SEVERE ¿V¿ OR TIGHT FOLDS) IN THE ROOM TEMPERATURE VULCANIZATION (RTV) SHELL OF SILTEX BREAST IMPLANTS. BREAST IMPLANTS ARE NOT CONSIDERED LIFETIME DEVICES AND DEFLATION IS A KNOWN COMPLICATION ASSOCIATED WITH THESE DEVICES AND IS REFERENCED IN OUR PRODUCT INSERT DATA SHEET. (B)(4).
THE MENTOR FAILURE ANALYSIS LAB HAS RECEIVED THE DEVICE FOR EVALUATION. THE ANALYSIS HAS BEGUN, BUT IS NOT COMPLETE AT THIS TIME. WHEN THE INVESTIGATIONAL ANALYSIS HAS BEEN COMPLETED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THE DEVICE HISTORY RECORD (DHR) WAS REVIEWED, AND NO ANOMALIES WERE FOUND RELATED TO THIS COMPLAINT. IN ADDITION, THE DHR REVIEW VERIFIES THAT THE DEVICE WAS MANUFACTURED IN ACCORDANCE WITH DOCUMENTED SPECIFICATION AND PROCEDURES. CONCOMITANT PRODUCTS: MENTOR SILTEX ROUND MODERATE PROFILE 250CC SALINE, CATALOG: 3542635, LOT: 6785720, SN: (B)(4). MANUFACTURER¿S REFERENCE NUMBER: (B)(4).
IT WAS REPORTED THAT A (B)(6) CAUCASIAN FEMALE PATIENT UNDERWENT BREAST AUGMENTATION REVISION WITH MENTOR SILTEX ROUND MODERATE PROFILE 250CC SALINE BREAST PROSTHESES. THE PATIENT REPORTED EXPERIENCING BURNING, TINGLING SENSATION AND VISIBLY DETERMINED THAT THE RIGHT BREAST PROSTHESIS HAD DEFLATED. AS A RESULT, THE PATIENT UNDERWENT BILATERAL REMOVAL AND REPLACEMENT WITH MENTOR SILTEX ROUND MODERATE PROFILE 250CC SALINE BREAST PROSTHESES ON (B)(6) 2018.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 834733 | MENTOR SILTEX ROUND MODERATE PROFILE | PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE | FWM | MENTOR TEXAS | 6861915 | 00081317001768 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 64 YR | Required Intervention |