FDA Adverse Event Malfunction Summary report: N

BD¿ 10 ML SYRINGE WITH INTERLINK® VIAL ACCESS

MDR report key: 7986812 · Received October 19, 2018

Report

Report Number
1213809-2018-00679
Event Type
Malfunction
Date Received
October 19, 2018
Date of Event
October 1, 2018
Report Date
January 9, 2019
Manufacturer
BECTON DICKINSON MEDICAL SYSTEMS
Product Code
FMI
UDI-DI
30382903034056
PMA / PMN Number
K924439
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION SUMMARY: ONE OPENED BLISTERPACK CONTAINING A 10ML SYRINGE WITH NEEDLE ATTACHED WAS RECEIVED IN AN ENVELOPE. THE PACKAGE WAS CONFIRMED TO BE FROM BATCH #8010900 (P/N 303405). STOPPER WAS OBSERVED TO HAVE A LARGE PIECE OF FOREIGN MATTER ATTACHED TO THE TOP, IN THE FLUID PATH. THE FOREIGN MATTER APPEARED TO BE A LARGE PIECE OF TAPE, LARGER THAN LEVEL 3 IN SIZE. POTENTIAL ROOT CAUSE FOR THE FOREIGN MATTER DEFECT IS ASSOCIATED WITH THE ASSEMBLY PROCESS. NO CORRECTIVE ACTIONS ARE NECESSARY BASED ON THE DEFECTIVE RATE IDENTIFIED. A DEVICE HISTORY RECORD REVIEW SHOWED NO REJECTED INSPECTIONS OR QUALITY ISSUES DURING THE PRODUCTION OF THE PROVIDED LOT NUMBER THAT COULD HAVE CONTRIBUTED TO THE REPORTED DEFECT.

Description of Event or Problem · 0

IT WAS REPORTED THAT BD¿ 10 ML SYRINGE WITH INTERLINK® VIAL ACCESS CONTAINED FOREIGN MATTER. NO SERIOUS INJURY OR MEDICAL INTERVENTION WAS REPORTED.

Additional Manufacturer Narrative · 1

DATE OF EVENT: UNKNOWN. A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED, BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT BD¿ 10 ML SYRINGE WITH INTERLINK® VIAL ACCESS CONTAINED FOREIGN MATTER. NO SERIOUS INJURY OR MEDICAL INTERVENTION WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
826000 BD¿ 10 ML SYRINGE WITH INTERLINK® VIAL ACCESS HYPODERMIC SYRINGE WITH NEEDLE FMI BECTON DICKINSON MEDICAL SYSTEMS 8010900 30382903034056

Patients

Seq Age Sex Outcome Treatment
1 Other