FDA Adverse Event Malfunction Summary report: N

BD POSIFLUSH¿ XS PRE-FILLED FLUSH SYRINGE

MDR report key: 7986587 · Received October 19, 2018

Report

Report Number
9616657-2018-00043
Event Type
Malfunction
Date Received
October 19, 2018
Date of Event
October 5, 2018
Report Date
November 2, 2018
Manufacturer
BECTON, DICKINSON AND CO.
Product Code
NGT
UDI-DI
00382903065721
PMA / PMN Number
N/A
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
PHARMACIST

Narratives

Additional Manufacturer Narrative · 0

ONE SAMPLE UNIT WAS RETURNED FOR EVALUATION BY OUR QUALITY ENGINEER. UPON VISUAL INSPECTION OF THE SAMPLE, THE BARREL LABEL WAS OBSERVED MISSING. THE BARREL LABEL PROVIDES THE SCALE MARKINGS AS WELL AS THE PRODUCT INFORMATION, INCLUDING THE LOT NUMBER AND EXPIRATION DATE. A DEVICE HISTORY RECORD REVIEW DID NOT REVEAL ANY DOCUMENTED QUALITY ISSUES DURING THE PRODUCTION OF LOT NUMBER 8143609 THAT COULD HAVE CONTRIBUTED TO THE OBSERVED DEFECT. BASED ON THE INVESTIGATION, IT HAS BEEN DETERMINED THAT THE MISSING BARREL LABELED WAS A RESULT OF A FAILURE IN THE VISION CAMERA. CORRECTIVE ACTION HAS BEEN INITIATED TO IMPROVE THE CAMERA DETECTION AND ENSURE DEFECTIVE PRODUCT IS DETECTED. COMPLAINTS FOR THIS DEFECT AND DEVICE WILL CONTINUE TO THE MONITORED.

Description of Event or Problem · 0

IT WAS REPORTED THAT BD POSIFLUSH XS PRE-FILLED FLUSH SYRINGES ARE MISSING SCALES ON THE SYRINGE. THERE WAS NO REPORT OF EXPOSURE, SERIOUS INJURY, OR MEDICAL INTERVENTION.

Additional Manufacturer Narrative · 1

DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD.(B)(6). A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT BD POSIFLUSH¿ XS PRE-FILLED FLUSH SYRINGES ARE MISSING SCALES ON THE SYRINGE. THERE WAS NO REPORT OF EXPOSURE, SERIOUS INJURY, OR MEDICAL INTERVENTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
828005 BD POSIFLUSH¿ XS PRE-FILLED FLUSH SYRINGE SALINE FLUSH NGT BECTON, DICKINSON AND CO. 8143609 00382903065721

Patients

Seq Age Sex Outcome Treatment
1 Other