FDA Adverse Event Malfunction Summary report: N

SYRINGES

MDR report key: 7985734 · Received October 19, 2018

Report

Report Number
1000113657-2018-01076
Event Type
Malfunction
Date Received
October 19, 2018
Date of Event
April 21, 2017
Report Date
October 19, 2018
Manufacturer
TRIVIDIA HEALTH, INC.
Product Code
NBW
PMA / PMN Number
K9913017
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). RETURNED PRODUCT EVALUATED WITH DEFECT FOUND. QC EVALUATION ON (B)(6) 2017 RESULT IN REPORTED DEFECT FOUND; INSPECTED THE SUBJECT SYRINGE AND DID FIND A RATHER LARGE PIECE OF UNKNOWN (LOOKS LIKE A PLASTIC FILM) DEBRIS IN THE NEEDLE AREA. MLC-65: MANUFACTURING OR PACKAGING DEFECT BASED ON THE POST MARKET REVIEW (B)(4) AND REVIEW OF THE FDA'S TOTAL PRODUCT LIFE CYCLE (TPLC) - TOTAL DEVICE PROBLEMS, FOR POSSIBLE DELAY IN ADMINISTRATION OF INSULIN. REFERENCE - CAPA (B)(4). NOTE: MANUFACTURER CONTACTED CUSTOMER (SEVERAL ATTEMPTS) IN A FOLLOW-UP CALL TO ENSURE THAT THE REPLACEMENT PRODUCTS RESOLVED THE INITIAL CONCERN - UNABLE TO ESTABLISH CONTACT WITH THE CUSTOMER AT THIS TIME. PRODUCT NOTIFICATION LETTER SENT TO CUSTOMER TO CONTACT CUSTOMER CARE.

Description of Event or Problem · 1

THE PHARMACIST CALLED AND STATED WHEN HE OPEN THE BOXES OF SYRINGES, IT APPEARS THAT THERE IS A PIECE OF PLASTIC COVERING THE NEEDLES. NO ONE GOT STUCK WITH THE NEEDLES. HE IS JUST CONCERN ABOUT THE APPEARANCE. THE NDC NUMBERS ARE: (B)(4) 29G 0.5CC HE HAS 2 AND A HALF BOXES (B)(4) 29G 1CC. HE HAS HALF BOX. I ADVISE THE PHARMACIST THAT WE WILL BE SENDING 4 BOXES OF THE GNP SYRINGES FOR HIM. HE ALREADY TRIED TO CALL HIS SUPPLIER AND WAS ADVISE TO CALL US. HE STATED THIS IS NOT THE FIRST TIME THIS HAS HAPPENED BUT HE HAD DISCARDED THE SYRINGES. I ADVISE HIM IN THE FUTURE TO CALL US AND NOT TO DISCARD THE PRODUCTS SINCE WE NEED IT FOR INVESTIGATION. I WILL REPLACE 4 BOX OF INSULIN NEEDLES AND HE WILL SEND THE OTHERS BACK FOR INVESTIGATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
826155 SYRINGES BLOOD GLUCOSE SYSTEM NBW TRIVIDIA HEALTH, INC. SYRINGES NA

Patients

Seq Age Sex Outcome Treatment
1 0 YR