BD INSYTE¿ AUTOGUARD¿ BC
Report
- Report Number
- 1710034-2018-00746
- Event Type
- Malfunction
- Date Received
- October 18, 2018
- Date of Event
- September 27, 2018
- Report Date
- December 14, 2018
- Manufacturer
- BECTON DICKINSON INFUSION THERAPY SYSTEMS INC.
- Product Code
- FOZ
- PMA / PMN Number
- N/A
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- BR
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
INVESTIGATION SUMMARY: ONLY A VIDEO WAS SUBMITTED FOR REVIEW. A COMPLAINT HISTORY CHECK WAS PERFORMED, AND THIS IS THE 1ST RELATED COMPLAINT REPORTED WITH THE DEFECT/CONDITION OF CATHETER DEFECTIVE / DAMAGED WITH LOT #8157541 REGARDING ITEM #38101214. DHR REVIEW WAS PERFORMED ON LOT NUMBER 8031566. THE LOT NUMBER WAS BUILT ON AFA LINE 10 FROM 08FEB2018 THRU 09FEB2018. REVIEW OF DHR REVEALED ALL REQUIRED CHALLENGE SAMPLES, SET-UP AND IN PROCESS TESTING WAS PERFORMED IN ACCORDANCE WITH THE QUALITY PLANS. REVIEW DISCLOSED NO REJECT ACTIVITY FINDINGS THROUGHOUT THE BUILD OF THIS LOT THAT WOULD IMPACT THE OUTCOME OF THE QUALITY OF THE PRODUCT RELEVANT TO THE DEFECT STATED IN THE PIR. NO QUALITY NOTIFICATION WERE INITIATED DURING THE BUILD OF THIS LOT NUMBER. ALTHOUGH THE PHOTO DISPLAYED THE CATHETER/ADAPTER DEFECTIVE, UNCERTAIN WHERE THE DEFECT OCCURRED. THE UNIT WAS NOT RETURNED FOR EVALUATION AND TESTING. WITHOUT THE ACTUAL SAMPLE FOR EVALUATION AND TESTING THERE WAS NO PHYSICAL EVIDENCE TO CONFIRM OR SUPPORT MANUFACTURING PROCESS RELATED ISSUES FOR THE REPORTED DEFECT. A FORMAL CORRECTIVE ACTION WILL NOT BE INITIATED AT THIS TIME. CUSTOMER COMPLAINT TRENDS ARE EVALUATED MONTHLY. IF THE TREND OF A SPECIFIC TYPE OF COMPLAINT WARRANTS A FORMAL CORRECTIVE ACTION, RESOURCES WILL BE ASSIGNED AT THAT TIME. THE PEURA (END USER RISK ANALYSIS) WAS ANALYZED TO DETERMINE THE RISK TO CUSTOMER. THE ANALYSIS SHOWED THAT DUE TO LOW OCCURRENCE, CURRENT RISK IS ACCEPTABLE.
IT WAS REPORTED THAT THE BD INSYTE¿ AUTOGUARD¿ BC HAD A HOLE BETWEEN THE HUB AND THE BEVEL. THERE WAS NO REPORT OF EXPOSURE, INJURY, OR MEDICAL INTERVENTION.
DATE OF EVENT UNK - THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. (B)(6). OTHER OCCUPATION: (B)(6) - INFORMATION AND SECURITY SERVICE OF MEDICINES. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
IT WAS REPORTED THAT THE BD INSYTE¿ AUTOGUARD¿ BC HAD A HOLE BETWEEN THE HUB AND THE BEVEL. THERE WAS NO REPORT OF EXPOSURE, INJURY, OR MEDICAL INTERVENTION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 817319 | BD INSYTE¿ AUTOGUARD¿ BC | INTRAVASCULAR CATHETER | FOZ | BECTON DICKINSON INFUSION THERAPY SYSTEMS INC. | 8157541 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |