FDA Adverse Event
Malfunction
Summary report: N
VITROS 250 CHEMISTRY SYSTEM
MDR report key: 797827
·
Received December 18, 2006
Report
- Report Number
- 1319681-2006-00284
- Event Type
- Malfunction
- Date Received
- December 18, 2006
- Date of Event
- November 28, 2006
- Report Date
- November 28, 2006
- Manufacturer
- ORTHO-CLINICAL DIAGNOSTICS, INC.
- Product Code
- JJE
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MS, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
INVESTIGATION INTO THIS EVENT DETERMINED THAT THERE WERE SALT DEPOSITS ON THE ELECTROLYTE REFERENCE FLUID (ERF) EVAPORATION COVER. CLEANING OF THE EVAPORATION COVER RESTORED THE SYSTEM TO EXPECTED OPERATION. THE ROOT CAUSE OF THE EVENT IS USER ERROR IN NOT CLEANING THE ERF EVAPORATION COVER AFTER SPILLING ERF DURING DAILY MAINTENANCE.
Description of Event or Problem · 1
A CUSTOMER OBSERVED POSITIVELY BIASED K+ QC RESULTS ON THE VITROS 250 ANALYZER. BIASED RESULTS OF THE MAGNITUDE AND DIRECTION OBSERVED MAY LEAD TO INAPPROPRIATE PHYSICIAN ACTION. THERE WAS NO REPORT OF PATIENT HARM AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | VITROS 250 CHEMISTRY SYSTEM | CLINICAL CHEMISTRY ANALYZER | JJE | ORTHO-CLINICAL DIAGNOSTICS, INC. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | * |