FDA Adverse Event Malfunction Summary report: N

AVALON 13 FRENCH CANNULA

MDR report key: 7973984 · Received October 17, 2018

Report

Report Number
8010762-2018-00281
Event Type
Malfunction
Date Received
October 17, 2018
Date of Event
September 24, 2018
Report Date
January 17, 2019
Manufacturer
MAQUET CARDIOPULMONARY AG
Product Code
DWF
PMA / PMN Number
K081820
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

(B)(4). A FOLLOW-UP MEDWATCH WILL BE SUBMITTED WHEN ADDITIONAL INFORMATION BECOMES AVAILABLE. REFERENCE EXEMPTION # E2018002. (B)(4).

Additional Manufacturer Narrative · 0

MAQUET MEDICAL SYSTEMS,USA(IMPORTER)SUBMITS THIS REPORT ON BEHALF OF THE LEGAL MANUFACTURER OF THE DEVICE MAQUET CARDIOPULMONARY AG KEHLER STRASSE 31, 76437 RASTATT, GERMANY. A FOLLOW-UP MEDWATCH WILL BE SUBMITTED WHEN ADDITIONAL INFORMATION BECOMES AVAILABLE. REFERENCE EXEMPTION # E2018002. IMPORTER- MAQUET MEDICAL SYSTEMS USA 45 BARBOUR POND DRIVE WAYNE, NJ 07470. CONTACT PERSON- (B)(6). THE DEFECTIVE PART HAS ARRIVED FOR INVESTIGATION: AN AVALON ELITE 13 FR HAS BEEN RETURNED. THE SAMPLE WAS CONTAMINATED. SAMPLE WAS CLEANED WITH SODIUM HYPOCHLORITE. VISUAL INSPECTION PERFORMED. NO ABNORMALITIES COULD BE DETECTED. THE INTRODUCER COULD BE PULLED OUT OF THE CATHETER NORMALLY AND PUSHED BACK IN AGAIN. A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED IF ADDITIONAL INFORMATION BECOMES AVAILABLE.

Description of Event or Problem · 0

ACCORDING TO THE CUSTOMER: IT WAS REPORTED THAT AFTER INSERTION OF THE 13 FRENCH AVALON CANNULA INTO THE RIGHT IJ THE OBTERATOR COULD NOT BE REMOVED. THE CANNULA WAS THEN REMOVED AND ECLS SUPPORT WAS INITIATED THROUGH CENTRAL CANNULATION. ADDITIONAL INFORMATION: THERE WERE NO NEGATIVE CONSEQUENCES FOR THE PATIENT REPORTED. (B)(4).

Description of Event or Problem · 0

REF.: (B)(4). CUSTOMER REF.: (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
817020 AVALON 13 FRENCH CANNULA CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS DWF MAQUET CARDIOPULMONARY AG

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention