FDA Adverse Event Malfunction Summary report: N

SERVO-U

MDR report key: 7973586 · Received October 17, 2018

Doing adverse event research by hand?

The Research Assistant searches MAUDE for you, groups reports by failure mode, and cites every record.

Free to try · every answer cites its records

Report

Report Number
8010042-2018-00548
Event Type
Malfunction
Date Received
October 17, 2018
Date of Event
September 28, 2018
Report Date
December 6, 2018
Manufacturer
MAQUET CRITICAL CARE AB
Product Code
CBK
PMA / PMN Number
K151814
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

GETINGE USA SALES, LLC (IMPORTER) IS SUBMITTING THIS REPORT ON BEHALF OF MAQUET CRITICAL CARE AB (MANUFACTURER). REF. EXEMPTION #: E2018003. GETINGE USA SALES, LLC 45 BARBOUR POND DRIVE WAYNE, NJ 07470. CONTACT PERSON: (B)(4).

Additional Manufacturer Narrative · 0

THE VENTILATOR WAS INVESTIGATED ON SITE AND THE O2 GAS MODULE WAS REPLACED AND RETURNED FOR MANUFACTURE INVESTIGATION. THE REPORTED FLUCTUATING O2 CONCENTRATION WERE REPRODUCED DURING TEST OF THE RETURNED O2 GAS MODULE MOUNTED IN A VENTILATOR. THE NOZZLE UNIT WAS FOUND TO CAUSING OSCILLATING IN THE O2 GAS MODULE. THE NOZZLE UNIT HAS A PREDETERMINED LIFETIME AND IT IS REPLACED AT PREVENTIVE MAINTENANCE (PM) THAT MUST BE PERFORMED EVERY 5000 HOURS OF OPERATION OR AT LEAST ONCE A YEAR. IF OSCILLATIONS HAPPEN DURING VENTILATION, THE PATIENT MAY RECEIVE AN AMOUNT OF OXYGEN IN THE GAS MIXTURE THAT DEVIATE FROM THE EXPECTED. FLOW AND PRESSURE MAY ALSO DEVIATE FROM THE EXPECTED. ALARMS WILL BE ACTIVATED IF THE FAILURE OCCURS DURING VENTILATION. GETINGE USA SALES, LLC (IMPORTER) IS SUBMITTING THIS REPORT ON BEHALF OF MAQUET CRITICAL CARE AB (MANUFACTURER). REF. EXEMPTION #: E2018003. GETINGE USA SALES, LLC (B)(4). CONTACT PERSON:(B)(4).

Description of Event or Problem · 0

IT WAS REPORTED THAT THE O2 CONCENTRATION WAS FLUCTUATING DURING PATIENT TREATMENT. THERE WAS NO PATIENT HARM. IMPORTER REF. #: (B)(4). MANUFACTURER REF. #: (B)(4).

Description of Event or Problem · 0

IMPORTER REF. #: (B)(4). MANUFACTURER REF. #: (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
816919 SERVO-U VENTILATOR, CONTINUOUS, FACILITY USE CBK MAQUET CRITICAL CARE AB

Patients

Seq Age Sex Outcome Treatment
1