LEEP PRECISION GENERATOR
Report
- Report Number
- 1216677-2018-00059
- Event Type
- Malfunction
- Date Received
- October 15, 2018
- Date of Event
- July 20, 2018
- Report Date
- October 15, 2018
- Manufacturer
- COOPERSURGICAL, INC.
- Product Code
- HGI
- PMA / PMN Number
- K963653
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
*INVESTIGATION X-REVIEW DHR X-INSPECT RETURNED SAMPLES *ANALYSIS AND FINDINGS A REVIEW OF THE 2 YR COMPLAINT HISTORY REVEALS SIMILAR ISSUES. A REVIEW OF THE DHR REVEALS NO ANOMALIES. SERVICE & REPAIR DID NOT CONFIRM THE COMPLAINT. NO ISSUES NOTED WITH THE UNIT'S FUNCTION OR THE FOOT PEDAL. THE UNIT OPERATED TO SPECIFICATIONS. A DEFINITIVE ROOT CAUSE IS NOT AVAILABLE FOR THIS COMPLAINT CONDITION. IT IS POSSIBLE THE FOOT PEDAL WAS IMPROPERLY USED. *CORRECTION AND/OR CORRECTIVE ACTION THE UNIT WAS UPDATED WITH A DIODE PER THE CURRENT BOARD REVISION AND RETURNED TO THE CUSTOMER. ALSO, THE FOOT PEDAL WAS REPLACED AS A PRECAUTION. THE CHANGE WAS EXECUTED VIA (B)(4). *PREVENTATIVE ACTION ACTIVITY COOPERSURGICAL WILL CONTINUE TO MONITOR THIS COMPLAINT CONDITION FOR ANY TRENDS.
FOOT PEDAL IS ONLY INTERMITTENTLY WORKING CURRENT IS COMING ON AND OFF BY ITSELF.
COOPER SURGICAL INC. IS CURRENTLY INVESTIGATING THE REPORTED COMPLAINT CONDITION. THE ACTUAL DEVICE INVOLVED IN THE COMPLAINT HAS BEEN RETURNED FOR EVALUATION BY THE CUSTOMER. ONCE THE INVESTIGATION IS COMPLETED A FOLLOW-UP REPORT WILL BE FILLED. (B)(4).
FOOT PEDAL IS ONLY INTERMITTENTLY WORKING. CURRENT IS COMING ON AND OFF BY ITSELF.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 804986 | LEEP PRECISION GENERATOR | LP-20-120 | HGI | COOPERSURGICAL, INC. | LP-20-120 | N/A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown | Other |