FDA Adverse Event Malfunction Summary report: N

LEEP PRECISION GENERATOR

MDR report key: 7966738 · Received October 15, 2018

Report

Report Number
1216677-2018-00059
Event Type
Malfunction
Date Received
October 15, 2018
Date of Event
July 20, 2018
Report Date
October 15, 2018
Manufacturer
COOPERSURGICAL, INC.
Product Code
HGI
PMA / PMN Number
K963653
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

*INVESTIGATION X-REVIEW DHR X-INSPECT RETURNED SAMPLES *ANALYSIS AND FINDINGS A REVIEW OF THE 2 YR COMPLAINT HISTORY REVEALS SIMILAR ISSUES. A REVIEW OF THE DHR REVEALS NO ANOMALIES. SERVICE & REPAIR DID NOT CONFIRM THE COMPLAINT. NO ISSUES NOTED WITH THE UNIT'S FUNCTION OR THE FOOT PEDAL. THE UNIT OPERATED TO SPECIFICATIONS. A DEFINITIVE ROOT CAUSE IS NOT AVAILABLE FOR THIS COMPLAINT CONDITION. IT IS POSSIBLE THE FOOT PEDAL WAS IMPROPERLY USED. *CORRECTION AND/OR CORRECTIVE ACTION THE UNIT WAS UPDATED WITH A DIODE PER THE CURRENT BOARD REVISION AND RETURNED TO THE CUSTOMER. ALSO, THE FOOT PEDAL WAS REPLACED AS A PRECAUTION. THE CHANGE WAS EXECUTED VIA (B)(4). *PREVENTATIVE ACTION ACTIVITY COOPERSURGICAL WILL CONTINUE TO MONITOR THIS COMPLAINT CONDITION FOR ANY TRENDS.

Description of Event or Problem · 0

FOOT PEDAL IS ONLY INTERMITTENTLY WORKING CURRENT IS COMING ON AND OFF BY ITSELF.

Additional Manufacturer Narrative · 1

COOPER SURGICAL INC. IS CURRENTLY INVESTIGATING THE REPORTED COMPLAINT CONDITION. THE ACTUAL DEVICE INVOLVED IN THE COMPLAINT HAS BEEN RETURNED FOR EVALUATION BY THE CUSTOMER. ONCE THE INVESTIGATION IS COMPLETED A FOLLOW-UP REPORT WILL BE FILLED. (B)(4).

Description of Event or Problem · 1

FOOT PEDAL IS ONLY INTERMITTENTLY WORKING. CURRENT IS COMING ON AND OFF BY ITSELF.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
804986 LEEP PRECISION GENERATOR LP-20-120 HGI COOPERSURGICAL, INC. LP-20-120 N/A

Patients

Seq Age Sex Outcome Treatment
1 Unknown Other