FDA Adverse Event Malfunction Summary report: N

MXAPP TRIMO SAN JEL-JECTOR

MDR report key: 7966334 · Received October 15, 2018

Report

Report Number
1216677-2018-00052
Event Type
Malfunction
Date Received
October 15, 2018
Date of Event
September 25, 2018
Report Date
October 15, 2018
Manufacturer
COOPERSURGICAL, INC.
Product Code
HGD
PMA / PMN Number
E216669
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

*INVESTIGATION NO SAMPLE RETURNED REVIEW DHR INSPECT STOCK PRODUCT *ANALYSIS AND FINDINGS AN EVALUATION OF THE COMPLAINANT SAMPLES COULD NOT BE CONDUCTED SINCE THEY WILL NOT BE RETURNED. THE APPLICATOR ASSEMBLY IS MADE UP OF THREE COMPONENT PARTS (THE BULB, THE FLYER AND THE TUBE) PURCHASED FROM AND ASSEMBLED BY A SUPPLIER. A REVIEW OF TWO YEAR COMPLAINT HISTORY SHOWS SOME COMPLAINTS FOR THIS ISSUE. A REVIEW OF THE DHR (243208) SHOW NO ABNORMALITIES. A REVIEW OF COOPERSURGICAL INCOMING INSPECTION RECORDS OF THE TUBE OVER THE PAST 2 YEARS SHOWS NO REJECTIONS FOR THIS ISSUE. THE TUBE IS MOLDED BY A SUPPLIER AND HAS 8 CAVITIES. A SAMPLE INSPECTION OF ALL 8 CAVITIES IN STOCK INVENTORY FOUND ALL THE PARTS MET QUALITY REQUIREMENTS AND WERE NOT SHARP AT THE DISTAL END. THE REPORTED COMPLAINT COULD NOT BE CONFIRMED. *CORRECTION AND/OR CORRECTIVE ACTION / *PREVENTATIVE ACTION ACTIVITY WILL CONTINUE TO MONITOR FOR TRENDING.

Description of Event or Problem · 0

"PATIENT WENT TO USE THE APPLICATOR AND THE END WAS SHARP AND SCRAPE/CUT HER ON THE INSIDE OF HER VAGINAL CANAL. " REFERENCE E-COMPLAINT (B)(4).

Additional Manufacturer Narrative · 1

(B)(4). THE COMPLAINT CONDITION REPORTED IS CURRENTLY BEING INVESTIGATED. ONCE THE INVESTIGATION IS COMPLETED A FOLLOW UP REPORTED WILL BE FILED.

Description of Event or Problem · 1

"PATIENT WENT TO USE THE APPLICATOR AND THE END WAS SHARP AND SCRAPE/CUT HER ON THE INSIDE OF HER VAGINAL CANAL. " (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
804993 MXAPP TRIMO SAN JEL-JECTOR TRIMO SAN APPLICATOR HGD COOPERSURGICAL, INC. MXAPP 6628

Patients

Seq Age Sex Outcome Treatment
1 Unknown Other