2.7 DEGREE STRAIGHT SAGITAL SAW

MDR report key: 7965500 · Received October 15, 2018

Report

Report Number
3005985723-2018-00588
Event Type
Malfunction
Date Received
October 15, 2018
Date of Event
October 9, 2018
Report Date
January 3, 2019
Manufacturer
MAKO SURGICAL CORP.
Product Code
OLO
UDI-DI
00848486032111
PMA / PMN Number
K143752
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

REPORTED ISSUE: UPON DISASSEMBLY OF MICS AT END OF CASE THERE WAS A SMALL METAL SHAVING OF UNKNOWN SOURCE ON ATTACHMENT. CASE TYPE: TKA. PRODUCT HISTORY REVIEW: A REVIEW OF THE DEVICE HISTORY RECORDS INDICATE 50 DEVICES WERE MANUFACTURED AND ACCEPTED INTO FINAL STOCK ON 09/16/17. NO NON-CONFORMANCES WERE IDENTIFIED DURING INSPECTION. COMPLAINT HISTORY REVIEW: A REVIEW OF COMPLAINTS RELATED TO P/N 212186, LOT NUMBER 35020817 SHOWS NO ADDITIONAL COMPLAINT(S) RELATED TO THE FAILURE IN THIS INVESTIGATION. TRACKING OF COMPLAINTS RELATED TO P/N 212186 WILL BE TRACKED THROUGH TREND REQUEST NUMBER (B)(4). VISUAL INSPECTION: VISUAL INSPECTION COULD NOT BE PERFORMED BECAUSE THE PRODUCT WAS NOT RETURNED. DIMENSIONAL INSPECTION: DIMENSIONAL INSPECTION COULD NOT BE PERFORMED BECAUSE THE PRODUCT WAS NOT RETURNED. MATERIAL ANALYSIS: MATERIAL ANALYSIS COULD NOT BE PERFORMED BECAUSE THE PRODUCT WAS NOT RETURNED. NC/CAPA HISTORY REVIEW: A REVIEW OF STRYKER¿S NC/CAPA DATABASE INDICATED THERE HAVE BEEN NO NCS OR CAPAS ASSOCIATED WITH THE PRODUCT AND FAILURE MODE REPORTED IN THIS EVENT. CONCLUSION: THE FAILURE MODE COULD NOT BE CONFIRMED BECAUSE THE PART WAS NOT RETURNED. IF DEVICE AND/OR ADDITIONAL INFORMATION BECOME AVAILABLE, THIS INVESTIGATION WILL BE REOPENED. DEVICE NOT RETURNED.

Description of Event or Problem · 0

UPON DISASSEMBLY OF MICS AT END OF CASE THERE WAS A SMALL METAL SHAVING OF UNKNOWN SOURCE ON ATTACHMENT. CASE TYPE: TKA.

Additional Manufacturer Narrative · 1

AS PART OF NORMAL COMPLAINT FOLLOW-UP, AN EVALUATION OF THE EVENT HAS BEEN INITIATED BY MAKO SURGICAL. A SUPPLEMENTAL REPORT WILL BE SUBMITTED WHEN ADDITIONAL INFORMATION BECOMES AVAILABLE.

Description of Event or Problem · 1

UPON DISASSEMBLY OF MICS AT END OF CASE THERE WAS A SMALL METAL SHAVING OF UNKNOWN SOURCE ON ATTACHMENT. CASE TYPE: TKA.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
806430 2.7 DEGREE STRAIGHT SAGITAL SAW STEREOTAXIC DEVICE, ROBOTICS OLO MAKO SURGICAL CORP. 35020817 00848486032111

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization