FDA Adverse Event Injury Summary report: N

DISTAL FEMUR JTS

MDR report key: 7964677 · Received October 15, 2018

Report

Report Number
3004105610-2018-00098
Event Type
Injury
Date Received
October 15, 2018
Date of Event
September 17, 2018
Report Date
March 20, 2019
Manufacturer
STANMORE IMPLANTS WORLDWIDE
Product Code
KRO
PMA / PMN Number
K133152
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 0

CORRECTED DATA: DEVICE NAME WAS CORRECTED FROM DISTAL FEMUR TO DISTAL FEMUR JTS. PMA 510(K) # WAS CORRECTED FROM K140898 TO K133152. ADDITIONAL MANUFACTURE NARRATIVE: AN EVENT REGARDING ALLEGED FRACTURE OF THE TIBIAL HINGE INVOLVING A JTS DISTAL FEMUR WAS REPORTED. THE EVENT WAS CONFIRMED BY MEDICAL REVIEW. METHOD AND RESULTS: PRODUCT EVALUATION AND RESULTS: NOT PERFORMED AS NO ITEMS WERE RETURNED. CLINICIAN REVIEW: THE IMPLANT IN SITU WAS FOR A JTS DISTAL FEMORAL REPLACEMENT WHICH WAS INSERTED ON (B)(6) 2011. THE SURGEON REPORTED THAT THE PATIENT HAD A FALL AND THE KNEE HINGE HAS BROKEN. THE MEDICAL IMAGING PROVIDED SHOWS THAT THE TIBIAL PART OF THE HINGE HAS FRACTURED, AND THE KNEE JOINT HAS DISLOCATED. THE FEMORAL STEM HAS MOVED ANTERIORLY, ESPECIALLY WHEN THE KNEE IS IN FLEXION. THE ABOVE RADIOGRAPHIC ASSESSMENT HAS CONFIRMED THE REASON FOR REVISION. PRODUCT HISTORY REVIEW: REVIEW OF THE PRODUCT HISTORY RECORDS INDICATE ONE DEVICE WAS MANUFACTURED AND ACCEPTED INTO FINAL STOCK ON 18 AUG 2011 WITH NO REPORTED DISCREPANCIES. COMPLAINT HISTORY REVIEW: THERE HAVE BEEN NO OTHER EVENTS. CONCLUSIONS: FOLLOWING A PATIENT FALL THE TIBIAL HINGE WAS REPORTED TO HAVE FRACTURED AND THE SURGEON REPORTED THAT THE AXLE CAP WAS REMOVED FROM ITS PLACE, THE SURGEON DOES NOT KNOW IF THE DISASSOCIATION OF THE AXLE CAP OCCURRED WHEN THE IMPLANT WAS BROKEN OR BEFORE. THE EXACT CAUSE OF THE EVENT COULD NOT BE DETERMINED BECAUSE FURTHER INFORMATION SUCH AS DEVICE RETURN, THE PRIMARY OPERATIVE REPORT AS WELL AS PATIENT HISTORY AND FOLLOW UP NOTES ARE NEEDED TO COMPLETE THE INVESTIGATION FOR DETERMINING ROOT CAUSE. NO FURTHER INVESTIGATION FOR THIS EVENT IS POSSIBLE AT THIS TIME AS NO DEVICES AND INSUFFICIENT INFORMATION WAS RECEIVED BY SIW. IF DEVICES AND ADDITIONAL INFORMATION BECOME AVAILABLE TO INDICATE FURTHER EVALUATION IS WARRANTED, THIS RECORD WILL BE RE-OPENED. REVISION SURGERY TOOK PLACE ON (B)(6) 2018 DUE TO FRACTURED TIBIAL HINGE WHEREBY JTS DISTAL FEMUR COMPONENTS WERE EXPLANTED. SIW WILL CONTINUE TO MONITOR FOR TRENDS. CORRECTIVE ACTION/PREVENTIVE ACTION: NO ACTION IS REQUIRED AT THIS TIME AS THERE IS NO INDICATION TO SUGGEST A PRODUCT NON-CONFORMITY OR UNANTICIPATED HAZARD.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE PATIENT WAS WALKING, LOST CONTROL OF HIS LEG, AND FELL. X-RAYS INDICATE THE HINGE OF THE PATIENT'S KNEE CONSTRUCT BROKE. A REVISION HAS NOT YET BEEN PERFORMED BUT IS BEING PLANNED TO EXCHANGE THE TIBIAL COMPONENT. FEMORAL COMPONENT REPORTED AS WELL-FIXED AND WILL NOT BE REVISED DUE TO ANY ALLEGATIONS AGAINST THE DEVICE.

Additional Manufacturer Narrative · 1

AN INVESTIGATION IS BEING PERFORMED IN AN ATTEMPT TO IDENTIFY THE CAUSE OF THE EVENT. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE IT WILL BE REPORTED IN A SUPPLEMENTAL REPORT. DEVICE NOT AVAILABLE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT WAS WALKING, LOST CONTROL OF HIS LEG, AND FELL. X-RAYS INDICATE THE HINGE OF THE PATIENT'S KNEE CONSTRUCT BROKE. A REVISION HAS NOT YET BEEN PERFORMED BUT IS BEING PLANNED TO EXCHANGE THE TIBIAL COMPONENT. FEMORAL COMPONENT REPORTED AS WELL-FIXED AND WILL NOT BE REVISED DUE TO ANY ALLEGATIONS AGAINST THE DEVICE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
808231 DISTAL FEMUR JTS LIMB SALVAGE SYSTEM KRO STANMORE IMPLANTS WORLDWIDE PIN 16331

Patients

Seq Age Sex Outcome Treatment
1 18 YR Hospitalization| R SMBPR00 SMILES KNEE BUMPERS EXSM B7775| SMBSH00 SMILES KNEE BUSHES EXSMALL B7776| SMCAP01 AXLE CAP B8537| SMTBC00G PASSIVE GROWER TIB BEARING XSM B8462