FDA Adverse Event Injury Summary report: N

ORA SYSTEM 2000

MDR report key: 7959881 · Received October 12, 2018

Report

Report Number
2028159-2018-02137
Event Type
Injury
Date Received
October 12, 2018
Report Date
December 11, 2018
Manufacturer
ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER
Product Code
NCF
PMA / PMN Number
EXEMPT
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION PROVIDED. THE COMPANY SERVICE REPRESENTATIVE EXAMINED THE SYSTEM AND WAS UNABLE TO REPLICATE THE REPORTED EVENT. NO ISSUES WERE IDENTIFIED WITH THE SYSTEM. THE SYSTEM WAS THEN TESTED AND MET ALL PRODUCT SPECIFICATIONS. BASED ON ASSESSMENT, THE PRODUCT MET SPECIFICATIONS AT THE TIME OF RELEASE. THE SYSTEM WAS FOUND TO MEET SPECIFICATIONS; THE ROOT CAUSE IS UNABLE TO BE DETERMINED CONCLUSIVELY. THE MANUFACTURER INTERNAL REFERENCE NUMBER IS: (B)(4).

Additional Manufacturer Narrative · 1

INVESTIGATION, INCLUDING ROOT CAUSE ANALYSIS, IS IN PROGRESS. A SUPPLEMENTAL MDR WILL BE FILED AS NECESSARY IN ACCORDANCE WITH 21 CFR 803.56 WHEN ADDITIONAL REPORTABLE INFORMATION BECOMES AVAILABLE. (B)(4).

Description of Event or Problem · 1

A CUSTOMER REPORTED FALSE READINGS DURING INTRAOPERATIVE ABERROMETRY. THE SITE REPORTS THIS HAS RESULTED IN TEN EXPLANTS DUE TO UNEXPECTED OUTCOMES. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
800099 ORA SYSTEM 2000 ABERROMETER, OPHTHALMIC NCF ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER NA ASKU

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention