FDA Adverse Event Malfunction Summary report: N

RESQPUMP

MDR report key: 7958524 · Received October 11, 2018

Report

Report Number
3003477173-2018-00005
Event Type
Malfunction
Date Received
October 11, 2018
Date of Event
September 18, 2018
Report Date
October 11, 2018
Product Code
PIZ
PMA / PMN Number
P110024
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
LA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE RESQPUMP ARRIVED WITH THE SUCTION CUP/CONNECTER STEM ASSEMBLY SEPARATED FROM THE HANDLE/SPRING ASSEMBLY. THE M5 WASHER WAS NOT INCLUDED IN THE RETURN HOWEVER THERE IS EVIDENCE THAT THERE HAD BEEN A WASHER INSTALLED PREVIOUSLY. WHEN OR HOW THE WASHER WAS SEPARATED FROM THE SYSTEM IS UNKNOWN. THERE WAS A GOOD AMOUNT OF WHITE PLASTIC SHAVINGS ON THE LOWER SPRING NUT AND INSIDE THE SUCTION CUP'S CONNECTION STEM RECESS. THE CLAMSHELL/STEM KEY FEATURE, WHICH ALIGNS THE TWO, SHOWED SIGNS OF DAMAGE AT THE TOP. THE DAMAGE IS PRESUMABLY DUE TO FRICTION WHILE THE M5 SCREW WAS IN THE PROCESS OF BACKING OUT. IT IS EVIDENT THAT THE M5 SCREW WAS ENGAGED IN THE THREADS (FULLY SEATED OR NOT IS UNKNOWN) AND BACKED OUT DURING OPERATION. BRASS SHAVINGS WERE IN THE SCREW THREADS ALONG THE ENTIRE ENGAGEMENT DEPTH INDICATING THAT, AT LEAST AT ONE TIME, THE SCREW WAS FULLY ENGAGED.

Description of Event or Problem · 1

THE RESQPUMP CAME APART DURING A RESUSCITATION EFFORT. AS FAR AS THE IMPACT IS UNKNOWN, MANUAL COMPRESSIONS WERE STARTED AND CONTINUED FROM THAT POINT FORWARD. I WAS NOT ON SCENE BUT AS FAR AS I CAN TELL, THE PT WAS BEING MOVED DOWN A FLIGHT OF STAIRS AND CPR BEING PERFORMED DURING THIS PROCESS. ONCE AT BOTTOM OF STAIRS THE DEVICE CAME APART LEAVING HALF ON PT CHEST AND OTHER HALF IN EMT'S HAND. IT SEEMS THAT THE NUT THAT HOLDS THE TWO TOGETHER CAME APART.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
797804 RESQPUMP RESQPUMP PIZ

Patients

Seq Age Sex Outcome Treatment
1