FDA Adverse Event
Other
Summary report: N
BD LOGIC BLOOD GLUCOSE MONITOR
MDR report key: 795763
·
Received December 22, 2004
Report
- Report Number
- 2243072-2004-00732
- Event Type
- Other
- Date Received
- December 22, 2004
- Date of Event
- November 27, 2004
- Report Date
- December 22, 2004
- Manufacturer
- BECTON DICKINSON
- Product Code
- NBW
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WA, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
AWAITING PRODUCT RETURN TO CONDUCT QUALITY INVESTIGATION.
Description of Event or Problem · 1
CONSUMER CALLED TO REPORT INCONSISTENT READINGS WITH HER BD LOGIC METER. ANALYSIS RUN ON CLARK ERROR GRID USING MEAN AVERAGE READING (373) AS REF POINT VS. BD READINGS (200, 545) YIELDED DATA POINTS IN THE C ZONE OF THE GRID, OUTSIDE ACCEPTABLE LIMITS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | BD LOGIC BLOOD GLUCOSE MONITOR | BD TEST STRIPS | NBW | BECTON DICKINSON | NA | 4064103 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNKNOWN | Other |