ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2018-18170
- Event Type
- Malfunction
- Date Received
- October 11, 2018
- Report Date
- October 1, 2018
- Manufacturer
- ANIMAS LLC
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- 003
Narratives
OF THE 79 ALLEGATIONS OF A MALFUNCTION, 48 PUMPS WERE RETURNED TO ANIMAS AND INVESTIGATED. OF THE INVESTIGATED PUMPS, 15 WERE FOUND TO BE MALFUNCTIONING DUE TO TRANSCEIVER CONNECTOR FAILURES.
IN QUARTER 4 OF 2018, THERE WERE 5 INSTANCES OF CS 052/053/054/055 SLEEP FAILURE FOUND DURING INVESTIGATION. THIS REPORT IS PROCESSED UNDER THE VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM.
THIS REPORT SUMMARIZES <NOE> 79</NOE> MALFUNCTION EVENTS. A REVIEW OF THE EVENTS INDICATED THAT THE ONE TOUCH PING MODEL PUMP EXPERIENCED A CS 052/053/054/055 SLEEP ISSUE. THESE REPORTS WERE RECEIVED FROM VARIOUS SOURCES. THE PATIENTS ASSOCIATED WITH THIS ALLEGATION RANGED FROM 4-74 YEARS OF AGE AND BETWEEN 15 AND 275 POUNDS. OF THE REPORTED PATIENTS, 35 WERE MALE, 35 WERE FEMALE, AND WERE UNKNOWN.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 795696 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS LLC |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |