FDA Adverse Event Malfunction Summary report: Y

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 7955476 · Received October 11, 2018

Report

Report Number
2531779-2018-18170
Event Type
Malfunction
Date Received
October 11, 2018
Report Date
October 1, 2018
Manufacturer
ANIMAS LLC
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 1

OF THE 79 ALLEGATIONS OF A MALFUNCTION, 48 PUMPS WERE RETURNED TO ANIMAS AND INVESTIGATED. OF THE INVESTIGATED PUMPS, 15 WERE FOUND TO BE MALFUNCTIONING DUE TO TRANSCEIVER CONNECTOR FAILURES.

Additional Manufacturer Narrative · 1

IN QUARTER 4 OF 2018, THERE WERE 5 INSTANCES OF CS 052/053/054/055 SLEEP FAILURE FOUND DURING INVESTIGATION. THIS REPORT IS PROCESSED UNDER THE VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM.

Description of Event or Problem · 1

THIS REPORT SUMMARIZES <NOE> 79</NOE> MALFUNCTION EVENTS. A REVIEW OF THE EVENTS INDICATED THAT THE ONE TOUCH PING MODEL PUMP EXPERIENCED A CS 052/053/054/055 SLEEP ISSUE. THESE REPORTS WERE RECEIVED FROM VARIOUS SOURCES. THE PATIENTS ASSOCIATED WITH THIS ALLEGATION RANGED FROM 4-74 YEARS OF AGE AND BETWEEN 15 AND 275 POUNDS. OF THE REPORTED PATIENTS, 35 WERE MALE, 35 WERE FEMALE, AND WERE UNKNOWN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
795696 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS LLC

Patients

Seq Age Sex Outcome Treatment
1