FDA Adverse Event Malfunction Summary report: N

PERMOBIL F5

MDR report key: 7953553 · Received October 10, 2018

Report

Report Number
1221084-2018-00065
Event Type
Malfunction
Date Received
October 10, 2018
Date of Event
September 13, 2018
Report Date
December 17, 2018
Manufacturer
PERMOBIL AB (PAB)
Product Code
ITI
PMA / PMN Number
K143014
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PO
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION REVEALED THAT THE ROOT CAUSE OF THIS FAILURE MODE IS LINKED TO A "COMPONENT FAILURE". THE MAIN WHEEL HUB BOLT ON THE DRIVE MOTOR SNAPPED CAUSING THE LEFT DRIVE TIRE TO DETACH. THE PATIENT WAS NOT HARMED. CLIENT INFORMATION WAS MISSING IN THE PREVIOUS REPORT (GENDER, WEIGHT) AND NOW SUPPLIED. THE SERVICING DEALER HAS BEEN SUPPLIED THE SPARE PARTS NEEDED TO PERFORM THE REPAIR AND CORRECT THIS ISSUE. THE WHEELCHAIR HAS BEEN RETURNED TO PATIENT OPERATING ACCORDING TO SPECIFICATION. THE DHR FOR THIS DEVICE HAS BEEN REVIEWED AND THE WHEELCHAIR MET SPECIFICATION PRIOR TO DISTRIBUTION.

Additional Manufacturer Narrative · 1

INVESTIGATION IS STILL ACTIVE. CONFIRMED REPORTS ARE THAT THE MAIN HUB BOLT BROKE AND ONE OF THE DRIVE WHEEL DETACHED. THIS IS A KNOWN ISSUE THAT HAS SINCE BEEN CORRECTED. THE FAILURE WAS LINKED BACK TO THE SUPPLIER AND RELATED TO AN ASSEMBLY ERROR INSTALLING THE WHEEL HUB TO THE DRIVE MOTOR SHAFT. THE RISK FOR THIS FAILURE TO OCCUR IS LOW AND THE LIKELY HOOD OF INJURY TO OCCUR IS ALSO VERY LOW. WE HAVE MADE MULTIPLE ATTEMPTS TO COLLECT MORE INFORMATION SUCH AS PATIENT GENDER, WEIGHT, STATUS OF THE WHEELCHAIR AND WHETHER THE NECESSARY SPARE PARTS HAVE BEEN SUPPLIED TO THE DEALER. ONCE COMMUNICATION CAN BE ESTABLISHED AND CORRECTIVE ACTION IMPLEMENTED A FOLLOW-UP REPORT WILL BE SUBMITTED. ANY INFORMATION CURRENTLY MISSING IN THIS REPORT WILL BE INCLUDED IN THE SUPPLEMENT. THE DHR FOR THIS DEVICE HAS BEEN REVIEWED AND THE WHEELCHAIR MET SPECIFICATION PRIOR TO DISTRIBUTION.

Description of Event or Problem · 1

REPORTED WHEEL BOLT BROKE WHILE PATIENT WAS DRIVING. NO INJURY HAS BEEN REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
794383 PERMOBIL F5 POWERED WHEELCHAIR ITI PERMOBIL AB (PAB) F5 N/A

Patients

Seq Age Sex Outcome Treatment
1