TRIAGE PROFILER SOB PANEL
Report
- Report Number
- 3013982035-2018-00027
- Event Type
- Malfunction
- Date Received
- October 5, 2018
- Report Date
- October 5, 2018
- Manufacturer
- QUIDEL CARDIOVASCULAR INC.
- Product Code
- MMI
- PMA / PMN Number
- K080269
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
PRODUCT DEVICE CODE (CONTINUED): DAP, DDR, AND JHX. INVESTIGATION CONCLUSION: THE CUSTOMER'S COMPLAINT WAS NOT REPLICATED DURING IN-HOUSE TESTING WITH RETAINS OF DEVICE LOT W64218B. NO ISSUES WITH TROPONIN I RECOVERY WERE OBSERVED; LOT PERFORMED PROPERLY. MANUFACTURING BATCH RECORDS FOR THE LOT WERE REVIEWED, LOT MET RELEASE SPECIFICATIONS. BASED ON THE INFORMATION AVAILABLE, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY AND NO CORRECTIVE ACTION IS REQUIRED. ALTHOUGH THIS CATALOG NUMBER (97300EU) IS NOT APPROVED IN THE UNITED STATES, THIS EVENT IS BEING REPORTED AS THE DEVICE IS SAME/SIMILAR TO CATALOG NUMBER 97300, 510(K) NUMBER K080269.
THE EVENT OCCURRED IN (B)(6). A PATIENT WHO PRESENTED WITH FLU SYMPTOMS AND PAIN IN CHEST AFTER WALKING 500 METERS WAS TESTED WITH THE TRIAGE SOB PANEL AND RECEIVED A TROPONIN I RESULT OF 0.06NG/ML. THE PATIENT WAS TRANSFERRED TO A SPECIALIST AND TESTED 30 MINUTES LATER WITH THE LABORATORY RADIOMETER AND RECEIVED A TROPONIN T RESULT OF 0.055NG/L. PATIENT WAS DIAGNOSED WITH A MYOCARDIAL INFARCTION AND HAD A STENT IMPLANTED. CUSTOMER STATED THERE WAS NO DELAY IN TREATMENT DUE TO THE TRIAGE RESULT. THE PATIENT IS NOW RELEASED FROM THE HOSPITAL AND IS STABLE. FACILITY TRIAGE TNI CUT-OFF: 0.4NG/ML. RADIOMETER TNT CUT-OFF: 0.02NG/L.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 781601 | TRIAGE PROFILER SOB PANEL | TRIAGE PROFILER SOB PANEL | MMI | QUIDEL CARDIOVASCULAR INC. | 97300EU | W64218B |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 59 YR |