FDA Adverse Event Malfunction Summary report: N

ATELLICA CH 930 ANALYZER

MDR report key: 7938297 · Received October 5, 2018

Report

Report Number
2432235-2018-00365
Event Type
Malfunction
Date Received
October 5, 2018
Date of Event
September 10, 2018
Report Date
October 16, 2018
Manufacturer
SIEMENS HEALTHCARE DIAGNOSTICS INC.
Product Code
JJE
UDI-DI
00630414002439
PMA / PMN Number
K151767
Removal / Correction Number
2432235-10/05/2018-003-C
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

SIEMENS FILED THE INITIAL MDR 2432235-2018-00365 ON 05-OCT-2018. CORRECTED INFORMATION (15-OCT-2018): THE INITIAL MDR INDICATED THAT THE SERUM INDICES CALCULATED WHEN SAMPLES ARE TESTED FOR DONOR METHODS [ALANINE AMINOTRANSFERASE (ALT), ASPARTATE AMINOTRANSFERASE (AST), LACTATE DEHYDROGENASE L- P (LDLP), AND UREA NITROGEN (UN_C)] ON THE ATELLICA SAMPLE HANDLER PRIME INSTRUMENT MAY NOT BE CONSISTENT WITH THE SAMPLES OR THE INDEPENDENTLY DERIVED H, I AND L INDICES OBTAINED ON THE INSTRUMENT. SIEMENS FURTHER INVESTIGATED THE ISSUE AND DETERMINED THAT THE INSTRUMENT IMPACTED BY THIS ISSUE WAS THE ATELLICA CH 930 ANALYZER.

Description of Event or Problem · 0

THE CUSTOMER REPORTED THAT THE SERUM INDICES (HEMOLYSIS [H], ICTERUS [I], AND LIPEMIA [L]) OBTAINED ON PATIENT SAMPLES ON THE ATELLICA CH 930 ANALYZER, WHEN THE SAMPLES WERE TESTED FOR UREA NITROGEN (UN_C) AND LACTATE DEHYDROGENASE L- P (LDLP), RECOVERED LESS THAN THE SERUM INDICES OBTAINED ON THE SAMPLES WHEN TESTED AT AN ALTERNATE LABORATORY. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE THE DIFFERENT SERUM INDICES OBTAINED ON THE PATIENT SAMPLES.

Additional Manufacturer Narrative · 1

SIEMENS HEALTHCARE DIAGNOSTICS HAS CONFIRMED THAT SERUM INDICES CALCULATED WHEN SAMPLES ARE TESTED FOR ALANINE AMINOTRANSFERASE (ALT), ASPARTATE AMINOTRANSFERASE (AST), LACTATE DEHYDROGENASE L- P (LDLP), AND UREA NITROGEN (UN_C) ON THE ATELLICA SAMPLE HANDLER PRIME INSTRUMENT MAY NOT BE CONSISTENT WITH THE SAMPLES OR THE INDEPENDENTLY DERIVED H, I, AND L INDICES OBTAINED ON THE INSTRUMENT. THE INVESTIGATION BY SIEMENS DETERMINED THAT THE ISSUE MAY CAUSE INCORRECT FLAGGING OF RESULTS, DEPENDING ON THE DIRECTION AND DEGREE OF INDEX DIFFERENCE AND THE ANALYTE HIL INDEX THRESHOLD. IN THE SCENARIO OF A HIGHER HIL INDEX THAN THE TRUE INDEX, RESULTS MAY BE FLAGGED FOR A HIGH SPECIMEN INDEX WHEN TRULY BELOW THE INDEX THRESHOLD FOR AN ANALYTE, WHICH WOULD BE APPARENT TO THE USER. IN THE SCENARIO OF A SIGNIFICANTLY LOWER HIL INDEX THAN THE TRUE INDEX, RESULTS MAY NOT BE FLAGGED AS APPROPRIATE. THE LIKELIHOOD OF INAPPROPRIATE FLAGGING LEADING TO AN INDEX CHANGE AND A SUBSEQUENT CLINICALLY SIGNIFICANT EFFECT IS REMOTE. SIEMENS IS NOT RECOMMENDING A REVIEW OF PREVIOUSLY GENERATED RESULTS. STARTING ON 26-SEP-2018, AN URGENT MEDICAL DEVICE CORRECTION (UMDC) ASW18-03.A.US WILL BE SENT TO US CUSTOMERS AND AN URGENT FIELD SAFETY NOTICE (UFSN) ASW18-03.A.OUS WILL BE SENT TO OUTSIDE THE US (OUS) CUSTOMERS. THE UDMC AND UFSN ADVISE CUSTOMERS TO NOT ORDER H, I, OR L SERUM INDICES IN THE SAME ORDER WITH ALT, AST, LDLP, OR UN_C TESTS. IN ORDER TO GET CORRECT SERUM INDICES VALUES, SIEMENS INSTRUCTED CUSTOMERS TO ORDER THE SERUM INDICES ON THE SYSTEM OR FROM THE LABORATORY INFORMATION SYSTEM (LIS), IN A SEPARATE ORDER WITHOUT ANY OF THE DONOR METHODS (ALT, AST, LDLP, OR UN_C). ALTERNATIVELY, THE CUSTOMER CAN TURN OFF THE AUTOMATIC HIL ORDERING FROM THE SETTINGS SCREEN: (SETUP->SETTINGS->GENERAL SETUP->PATIENT->HIL ORDERING) BY SELECTING "NEVER AUTOMATICALLY ORDER".

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SERUM INDICES (HEMOLYSIS [H], ICTERUS [I], AND LIPEMIA [L]) OBTAINED ON PATIENT SAMPLES ON THE ATELLICA SAMPLE HANDLER PRIME INSTRUMENT, WHEN THE SAMPLES WERE TESTED FOR UREA NITROGEN (UN_C) AND LACTATE DEHYDROGENASE L- P (LDLP), RECOVERED LESS THAN THE SERUM INDICES OBTAINED ON THE SAMPLES WHEN TESTED AT AN ALTERNATE LABORATORY. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE THE DIFFERENT SERUM INDICES OBTAINED ON THE PATIENT SAMPLES.