PERMOBIL C500
Report
- Report Number
- 1221084-2018-00061
- Event Type
- Injury
- Date Received
- October 5, 2018
- Date of Event
- December 3, 2017
- Report Date
- October 5, 2018
- Manufacturer
- PERMOBIL AB (PAB)
- Product Code
- ITI
- PMA / PMN Number
- K991658
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SW
- Reporter Occupation
- 003
Narratives
REPORTS CLAIM THE CHAIR ROLLED UNCONTROLLABLY DOWN A RAMP TO WHERE AT THE BOTTOM THE CHAIR TIPPED CAUSING THE END-USER LOSE POSITIONING IN THE SEATING AND FALL TO THE GROUND SUFFERING INJURIES TO INCLUDE A FRACTURED LEFT TIBIA AND FIBULA. REPORTS ARE THE END-USER WAS NOT WEARING THE INSTALLED POSITIONING BELT AT THE TIME OF THE EVENT. THE DEVICE WAS RETURNED TO PERMOBIL (B)(4) FOR EVALUATION. OPERATIONAL TESTING OF DEVICE ON AN 8° INCLINE SHOWN THE DEVICE TO ROLL APPROXIMATELY 6-8" BEFORE COMING TO A STOP AFTER THE JOYSTICK WAS RELEASED. THE REPORTS OF THE DEVICE CONTINUING TO ROLL UNCONTROLLABLY WHILE ON AN INCLINE WAS UNABLE TO BE DUPLICATED. THE DEVICE STOPPED AND HELD UNDER TESTING WITH RAMP ANGLES OF THE SAME DEGREE AS EVENT SITE OR GREATER. TESTING OF THE DEVICE INDICATED POTENTIAL MOTOR WEAR, BUT IS NOT BEING CONSIDERED AS EXCESSIVE FOR SAFE USE. MOTOR COMPENSATIONS AS WELL AS PARAMETER SETTINGS CAN BE ADJUSTED TO DECREASE THE AMOUNT OF TIME (ROLL) FOR COMPLETE STOPPAGE OF THE DEVICE. DUE TO THE EVENT HAVING OCCURRED OUT OF DOORS IN THE MONTH OF (B)(6), ENVIRONMENTAL FACTORS WERE NOT ABLE TO BE REPRODUCED, BUT WERE TAKEN INTO CONSIDERATION AS BEING A POSSIBLE CONTRIBUTING FACTOR. THE DHR WAS REVIEWED AND DEVICE MET SPECIFICATION PRIOR TO DISTRIBUTION.
RECEIVED REPORT STATING END-USER WAS IN THEIR WHEELCHAIR ATTEMPTING TO ENTER A WHEELCHAIR ACCESSIBLE MOTOR VEHICLE. IT WAS REPORTED THAT DURING THIS TRANSITION, THE WHEELCHAIR ALLEGEDLY ROLLED TO THE BOTTOM OF THE RAMP, UNCONTROLLABLY, TO WHERE IT STARTED TO TIP CAUSING THE END-USER TO LOSE POSITIONING IN THE SEATING AND FALL TO THE GROUND TO WHERE INJURIES WERE SUSTAINED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 779871 | PERMOBIL C500 | POWERED WHEELCHAIR | ITI | PERMOBIL AB (PAB) | C500 | N/A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Hospitalization |