LIFE PULSE HIGH FREQUENCY VENTILATOR
Report
- Report Number
- 1719232-2006-00004
- Event Type
- Other
- Date Received
- November 22, 2006
- Report Date
- November 22, 2006
- Manufacturer
- BUNNELL, INC.
- Product Code
- LSZ
- PMA / PMN Number
- p850064
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- OTHER
Narratives
REPORTED SYMPTOM OF "VENT FAULT" OR DEVICE SHUT OFF COLD NOT BE REPRODUCED. DEVICE POWERED UP WITH NO PROBLEMS. THE 6000 HOUR PREVENTIVE MAINTENANCE PERFORMED WITH NO FAILED COMPONENTS OR DEVICE MALFUNCTION. DEVICE CHECKED FOR SENSITIVITY TO AC LINE VOLTAGE. DEVICE OPERATED TO BELOW 105 VAC WITHOUT RESETTING OR SHUTTING DOWN. DEVICE FOUND TO PERFORM ACCORDING TO SPECIFICATIONS.
THE UNINTERRUPTIBLE POWER SOURCE (UPS) THE VENTILATOR WAS CONNECTED TO ALARMED FIRST. THEN THE LIFE PULSE VENTILATOR ALARMED "VENTILATOR FAULT" AND SHUT OFF (WENT INTO STANDBY MODE). THEY WERE NOT ABLE TO GET THE VENTILATOR BACK ON. THIS IS THE APPROPRIATE RESPONSE OF THE VENTILATOR TO A FLUCTUATION IN VOLTAGE. AFTER THE VENTILATOR FAULT ALARM, THE VENTILATOR HAS TO BE OFF FOR 3 MINUTES TO CLEAR THE MEMORY. ONCE THE MEMORY IS CLEARED, THE VENTILATOR WILL RESTART NORMALLY. THIS IS INCLUDED IN THE INSTRUCTIONS FOR USE. THE PATIENT WAS ON NITRIC OXIDE AT THE TIME OF THE INCIDENT. THE VENTILATOR WAS SWITCHED OUT WITH ANOTHER LIFE PULSE VENTILATOR. DURING THAT TIME, THE PATIENT WAS SUPPORTED ON MANUAL VENTILATION. DURING THAT TIME, THE PATIENT ARRESTED AND DIED. HAD THE CLINICIANS LEFT THE 1ST VENTILATOR OFF FOR THE REQUIRED 3 MINUTES AND RESTARTED IT, THE TIME THE PATIENT WAS OFF THE VENTILATOR MAY HAVE BEEN SHORTER. THE 24-HOUR HOT LINE TELEPHONE NUMBER IS AFFIXED TO THE VENTILATOR IN A VERY OBVIOUS WAY BUT NO CALL WAS RECEIVED FROM THE HOSPITAL. THE UPS CONNECTED TO THE VENTILATOR WAS NOT THE ONE SUPPLIED BY OR RECOMMENDED BY BUNNELL INC.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIFE PULSE HIGH FREQUENCY VENTILATOR | VENTILATOR | LSZ | BUNNELL, INC. | 203 | * |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | * | Other |