FDA Adverse Event
Malfunction
Summary report: N
ARCHITECT FSH CALIBRATOR
MDR report key: 793635
·
Received December 1, 2006
Report
- Report Number
- 2623532-2006-00025
- Event Type
- Malfunction
- Date Received
- December 1, 2006
- Date of Event
- July 24, 2006
- Report Date
- August 31, 2006
- Manufacturer
- ABBOTT DIAGNOSTICS INTERNATIONAL, LTD.
- Product Code
- CGJ
- PMA / PMN Number
- K032458
- Removal / Correction Number
- Z-0024-2007
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AS
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
FOLLOWING A CUSTOMER COMPLAINT, ABBOTT FOUND AN ATYPICAL STABILITY PROFILE OF ARCHITECT FSH CALIBRATOR LOT 33363Q100. THE INVESTIGATION TO DATE HAS SHOWN THAT BOTH CONTROLS AND PT RESULTS HAVE SHIFTED UPWARDS TOGETHER OVER TIME. THERE HAS BEEN NO ADVERSE EVENTS ASSOCIATED WITH THIS ISSUE TO DATE. THIS IS AN INTIAL REPORT. AN INVESTIGATION IS IN PROCESS. A FINAL REPORT WILL BE SUBMITTED WHEN THE INVESTIGATION IS COMPLETE.
Description of Event or Problem · 1
THE CUSTOMER STATES THAT THEY NOTICED A SHIFT UP IN CONTROLS (BIORAD) AND PT RESULTS WHEN USING THE ARCHITECT FSR ASSAY WITH ARCHITECT FSH CALIBRATORS LOT NUMBER 33363Q1. PT RESULTS WERE REPORTED OUT OF THE LAB WITH QC OUT OF RANGE. NO IMPACT TO PT MGMT WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ARCHITECT FSH CALIBRATOR | CMIA FOR THE QUANTITATIVE DETERMINATION OF FSH IN HUMAN SERU | CGJ | ABBOTT DIAGNOSTICS INTERNATIONAL, LTD. | NA | 33363Q100 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNKNOWN | ARCHITECT I2000 LIST NO. 8C89-01 |