MENTOR SMOOTH ROUND MODERATE PROFILE
Report
- Report Number
- 1645337-2018-05962
- Event Type
- Injury
- Date Received
- October 3, 2018
- Date of Event
- September 1, 2018
- Report Date
- September 14, 2018
- Manufacturer
- MENTOR TEXAS
- Product Code
- FWM
- UDI-DI
- 00081317001287
- PMA / PMN Number
- P990075
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
UPON RECEIPT BY MENTOR, THE DEVICE RETURNED WITHOUT FLUID. NO FOREIGN MATERIAL WAS OBSERVED WITHIN THE DEVICE. WHITE MATERIAL WAS OBSERVED ON THE SHELL SURFACE. UPON VISUAL EXAMINATION, THE DEVICE APPEARS INTACT. LEAK TESTING WAS PERFORMED IN ACCORDANCE WITH MENTOR PROCEDURES AND NO LEAK SITES WERE DETECTED. NO OTHER ANOMALIES WERE DISCOVERED. COMPLAINT WAS NOT CONFIRMED. THE DEVICE HISTORY RECORD (DHR) OF LOT NUMBER 5999818 WAS REVIEWED AND NO ANOMALIES WERE FOUND RELATED TO THIS COMPLAINT. IN ADDITION, THE DHR REVIEW VERIFIES THAT THE DEVICE WAS MANUFACTURED IN ACCORDANCE WITH DOCUMENTED SPECIFICATION AND PROCEDURES. THERE IS NO EVIDENCE THAT THE ISSUE IS RELATED WITH MANUFACTURING. BREAST IMPLANTS ARE NOT CONSIDERED LIFETIME DEVICES AND DEFLATION IS A KNOWN COMPLICATION ASSOCIATED WITH THESE DEVICES AND IS REFERENCED IN OUR PRODUCT INSERT DATA SHEET. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).
THE MENTOR FAILURE ANALYSIS LAB HAS RECEIVED THE DEVICE FOR EVALUATION. THE ANALYSIS HAS BEGUN, BUT IS NOT COMPLETE AT THIS TIME. WHEN THE INVESTIGATIONAL ANALYSIS HAS BEEN COMPLETED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THE DEVICE HISTORY RECORD (DHR) WAS REVIEWED, AND NO ANOMALIES WERE FOUND RELATED TO THIS COMPLAINT. IN ADDITION, THE DHR REVIEW VERIFIES THAT THE DEVICE WAS MANUFACTURED IN ACCORDANCE WITH DOCUMENTED SPECIFICATION AND PROCEDURES. CONCOMITANT MEDICAL PRODUCTS: MENTOR SMOOTH ROUND MODERATE PROFILE 425CC SALINE CATALOG: 3501670 LOT: 6001731 SN: (B)(4). MANUFACTURER¿S REFERENCE NUMBER: (B)(4).
IT WAS REPORTED THAT A (B)(6) CAUCASIAN FEMALE PATIENT UNDERWENT PRIMARY BREAST AUGMENTATION WITH MENTOR SMOOTH ROUND MODERATE PROFILE 425CC SALINE BREAST PROSTHESES. DEFLATION OF THE RIGHT BREAST PROSTHESIS WAS IDENTIFIED WHEN THE PATIENT WOKE UP. AS A RESULT, THE PATIENT UNDERWENT BILATERAL REMOVAL AND REPLACEMENT WITH MENTOR MEMORYGEL BREAST IMPLANT 600CC GEL BREAST PROSTHESES ON (B)(6) 2018.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 770458 | MENTOR SMOOTH ROUND MODERATE PROFILE | PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE | FWM | MENTOR TEXAS | 5999818 | 00081317001287 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 35 YR | Required Intervention |